TFA 31.5000 Handleiding

TFA Rookmelder 31.5000

Lees hieronder de 📖 handleiding in het Nederlandse voor TFA 31.5000 (4 pagina's) in de categorie Rookmelder. Deze handleiding was nuttig voor 28 personen en werd door 2 gebruikers gemiddeld met 4.5 sterren beoordeeld

Pagina 1/4
AIRCO2NTROL 3000 Manual
AIRCO2NTROL 3000 is a portable, accurate handheld CO2 monitor. In addition to measuring the CO2 concentration, this device can also
measure the ambient temperature. It uses NDIR technology to improve the long term stability.
Features:
VThe built-in Data logger can store 48 sets of CO2 and temperature in the past 24 hours; one log per 30 minutes.
VMax/Min mode can record the maximum and minimum concentration of CO2 since the device has been last turned on.
VThe Alarm mode will sound at 1000 PPM, the ASHRAE limit. The Alarm can be turned off.
VThe RCFS Mode can recover the original factory settings after the CO2 device has been recalibrated, altered, or damaged.
MODE FUNCTIONS
There are several Modes which we can adjust the setting parameters. These modes are Alti Mode,Alarm Mode, Outside Mode, Ucal Mode,
Datalogger Mode, MaxMin Mode and rcFS Mode in sequence.
ALTI Compensate the pressure changes with appropriate altitude of location when measure.
ALARM Alarm while CO2 concentration exceed one threshold, Alarm sound can be on or off.
OUTSIDE Modify the outside CO2 concentration, for defining the ventilation rate.
CALI Calibrate the sensor while the reading deviates from the actual CO2 concentration.
DATALOGGER Show the past CO2 and Temperature records in the past 24 hours.
MaxMin Show the Max and Min CO2 reading before being cleared or powered off after PowerOn.
rcFS Recover the factory setting, if the device cannot show accurate reading.
OPERATION INSTRUCTIONS
PowerOn
Press Power button() to turn on the device after placing 4 AA batteries into the battery holder or AC adapter plug into the DC socket .
Once power on, CO2 reading will show after 15 seconds of warm up.
1. Power Button 8. Function Label
2. Enter Button 9.Gas Entry Hole
3. Down Button 10. Power inlet
4. Up Button 11. RJ45 socket
5. Mode Button 12. Housing Stand
6. LCD Display 13. Battery Cover
7. LED Light 14. AA Batteries*4
(Display Features and Modes)
WarmUp
It lasts approximately 1min before WARMUP disappears; all the functions will not response during warm up.
UserMode
After warm up, the device will stabilize and display the normal CO 2 Reading (Upper Display)- remain visible at all times.
Temp and Ventilation Rates (Lower Display)
The Up/Down button(④ ③/ ) allows you to scroll through the Temperature and Ventilation Modes.When pressing the Up button,the
lower display will go through the following sequence: TempºC -> TempºF -> Vent Rate lps -> Vent Rate cfm/p
*Note: lps refers to Liter Per Second Per Person; cfm/p refers to Cubic Feet Per Minute Per Person
Operation of Mode Adjustment
1.ALTI Mode:
1.1. Press the Mode Button(), ALTI flashs
1.2. Press Mode() to alter between m(meters) and ft(feet) .
1.3. Press Up/Down(④ ③/ ) to adjust the altitude (Step=100m/500ft)
1.4. Press the Enter Button( ), save or leave the ALTI Mode, return to UserMode
2.ALARM Mode
2.1. Adjust the alarm level
2.1.1. Press the Mode button (), until ALARM flashs
2.1.2. Press Enter Button(), “CO2” icon flashs.
2.1.3. Press Up/Down(④ ③/ ) to adjust the alarm level (1,000 ppm, interval is ±100ppm; 1,000 ppm, interval is ±50 ppm)
2.1.4. Press the Enter(), save the setting and return to UserMode.
2.2. Turn ON/OFF the ALARM
2.2.1. Press the Mode button(), until “Speaker Icon” flashs.
2.2.2. Press Enter Button()
2.2.3. Press Up/Down(④ ③/ ) to turn on/off the ALARM,
2.2.4. Press the Enter(), save the setting and return to UserMode.
3.OUTSIDE Mode
** Note:Ventilation Rate
Ventilate rate represents how much air is introduced into the indoor space from the outside. Low values indicate low ventilation rates and
potentially poor air quality. High levels indicate excessive ventilation and potential excessive energy usage. To obtain an accurate
measurement, reading should be taken 2~3 hours after occupancy has stabilized in a space or at a peak in daily CO 2 concentrations.
In indoor air quality control, CO2 value is an indicator of ventilation rate. 400ppm (Parts Per Million) is the default CO2 concentration
outside (according to ASHRAE: American Society of Heating, Refrigeration and Air conditioning Engineers)
3.1. Press the Mode button() ,until OUTSIDE flashs
3.2. Press Enter(), show OUTSIDE, CO2 and PPM flash
Press Up/Down(④ ③/ ) to adjust the reading
3.3. Press the Enter(), save the setting and return to UserMode
After modification, the VENT Rate will change
4. CALI Mode
The CO2 monitor has been calibrated in the factory and should recalibrate every 12 months with a specified concentration of CO2.Sensor drift
usually occurs in linearity. Please follow below steps.
4.1. If battery icon shows low power, please replace new batteries or use AC adapter
4.2. Press the Mode button(), until CALI flashs
4.3. Press Enter(), CALI shows on display
4.4. Adjust the lower display to ambient CO 2 value by Up/Down (④ ③/ ) Button.
4.5. Press Mode button() more than 10 sec., CALIBRATING flashs.
Calibration will be done after 5 min and LCD will appear “PASS”.If LCD appear “FAIL”, please calibrate once again.
4.6. Press Enter(), return to UserMode
5. DATA LOGGER Mode
AIRCO2NTROL 3000 has a built-in datalogger , which can provide users the past CO2 and temperature readings within the past 24 hours.
5.1. Press the Mode button(), until DATALOGGER flashs
5.2. Press Enter(),CO2 and Temperature show up
Press Up/Down(④ ③/ ) to page up/down the reading .
5.3. Press the Enter(), return to UserMode
6. MaxMin Mode
AIRCO2NTROL 3000 has a built-in MaxMin mode, which can provide users the Maximum and Minimum CO2 readings since the device has
last been turned on.
6.1. Press the Mode button(), until MAX MIN flashs
6.2. Press Enter button(), MAX and MIN CO 2 reading alternatively shows.
6.3. “CLR” will flash if press Up/Down(④ ③/ )
Press Enter(),to CLEAR the MAX and MIN record .
6.4. Press Mode or Enter(), return to UserMode
7. rcFSMode:
If user set or calibrated the sensor wrongly, you can recover the factory setting
7.1. Press the Mode button(), until rcFS flashs on the lower display
7.2. Press Enter(), “no” shows on the upper display
7.3. Press Up/Down(④ ③/ ), “no” change to “yes”
Press Enter(), Factory Setting will reload, return to UserMode;
7.4. Press Mode(), leave rcFS Mode, return to UserMode
SPECIFICATIONS (Preliminary Version)
Method - NDIR
Display - LCD
Independent CO2 and Temperature readings.
Calculates and Displays Ventilation Rates
Sample Method -Diffusion or flow through (50 ~200 ml/min)
Performance - CO2
Measurement Range 0-3,000 ppm display
Display Resolution 1ppm at 0~1,000ppm; 5ppm at 1,000~2,000ppm; 10ppm at 2,000~3,000ppm
Accuracy ±50 ppm or ±5% of reading
Repeatability ±20 ppm
Temperature Dependence ±0.1% of reading per °C or ±2 ppm per °C, whichever is greater, referenced to 25°C
Pressure Dependence 0.13% of reading per mm Hg (Corrected via user input for altitude)
Response Time <2min for 63% of step change
Warm-Up Time <60 seconds at 22°C
Calibration Interval 12 months, offset adjustment using single gas at 0-1000 ppm CO2. Full factory calibration available
Performance - Temperature
Temperature Range Display 32 to 122°F (0 to 50°C)
Display Resolution 0.1°F (0.1°C)
Display Options °F / °C, or Off. Set with Up / Down button
Accuracy ±2°F (±1°C)
Response Time 20-30 minutes (case must equilibrate with environment)
Outputs
OC: Normally Low,100mA max. @ 24 VDC. Adjustable setpoint, factory setting is 1000ppm,hysteresis is 50ppm
One RJ-45 Connector digital output
Power Supply
Two power supply types
Battery type: Alkaline, AA* 4 (80 hours)
External: 6 VDC from external AC/DC adapter which is included in package(Use specified AC adapter only)
Power Requirement
t
t
tt
160 mA Peak, 15 mA average from 6V
General Operating Conditions
Operating Temperature: 32-122°F (0-50°C) 0-95% RH, non-condensing
Storage Temperature: -4 to 140°F (-20 to 60°C)
Dimensions
24.3×81.1×165 mm (0.96×3.19×6.50 inch)
Weight (including batteries)
273.6 grams(9.65 oz)
LCD ERROR MESSAGES
The device incorporates visual diagnostic messages as follows:
「 」 Error 3, it is displayed when the temperature being measured is outside of the measurement range.
「 」
「 」
Error 5~6, it is necessary to reset the thermometer. To reset it, turn the instrument off, remove the batter
y
and wait for
a minimum of one minute, reinsert the batter
y
and turn on. If the error messa
g
e remains please contact the Service
Department for further assistance.


Product specificaties

Merk: TFA
Categorie: Rookmelder
Model: 31.5000

Heb je hulp nodig?

Als je hulp nodig hebt met TFA 31.5000 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden




Handleiding Rookmelder TFA

Handleiding Rookmelder

Nieuwste handleidingen voor Rookmelder

Kidde

Kidde 3SFWR Handleiding

24 September 2024
Kidde

Kidde KF30R Handleiding

24 September 2024
Kidde

Kidde P4010ACS-WCA Handleiding

24 September 2024
Kidde

Kidde i12020CA Handleiding

24 September 2024
Kidde

Kidde i12040CA Handleiding

24 September 2024
Kidde

Kidde KF20R Handleiding

24 September 2024
Kidde

Kidde CP9000CA Handleiding

24 September 2024
Kidde

Kidde Pi2010CA Handleiding

24 September 2024