Operating Instructions
METRAHITWORLD
International TRMS Multimeter 3-349-529-03
10/6.17
2 GMC-I Messtechnik GmbH
1Display (LCD)
2ON / OFF
Light
Key for ON / OFF and display illumination
ON / OFF
3HOLD Key for the following functions: save
measured value, delete and MAX-MIN
4MAN / AUTO Key for manual measuring range selection
5FUNC Key for selecting ranges or functions
6rotary selector switch for measuring functions
7 Connector jacks
Symbols used in the Digital Display
1 MAN: manual measuring range selection is active
2 MAX/MIN value storage
3 HOLD: display memory, “freeze” measured value
4 Digital display with decimal point and polarity display
5 Diode measurement selected
6 Display Duty Cycle Measurement
7 appears when acoustic signal is active
8 Selected type of current
9 interface active
10 Unit of measure
11 Unit of measure C / F for temperature measurement
12 Rotation Per Minute measurement selected
13 Triangle appears: indicates overranging
14 Pointer for analog display
15 REL: relative measurement with reference to offset
16 Scale for analog display
17 Indicates that the negative analog display range has been exceeded
18 Warning regarding dangerous voltage: > 40 V AC / 60 DC
19 Display in case of defective fuse
20 Low battery display
21 Symbol for continuous duty
Standard Equipment
1 TRMS-digital multimeter
1 Protective rubber holster
2 1.5 V AA size batteries
1 Set of measurement cables KS17-2
1 DAkkS calibration certificate
1 Short-form operating instructions
1
2
3
4
5
6
7
Max.
1000 V !
14
542
12
13
9
1516
19
18
21
71 3
10
17
20
8
6
11
GMC-I Messtechnik GmbH 3
Table of Contents
Page Page
1 Safety Features and Precautions ...............................3
2 Initial Start-Up ............................................................5
3 Selecting Measuring Functions and Measuring
Ranges .......................................................................5
3.1 Automatic Measuring Range Selection ..................................... 5
3.2 Manual Measuring Range Selection ......................................... 5
3.3 Quick Measurements .............................................................. 5
3.4 Relative Measurement D REL ................................................... 6
4 LCD .............................................................................6
4.1 Digital Display .........................................................................6
4.2 Analog Display ........................................................................ 6
4.3 Display Illumination ................................................................. 6
5 Measured Value Memory “HOLD” ..............................6
6 Saving Minimum or Maximum Values
“MIN/MAX” Hold .........................................................6
7 Voltage Measurement ................................................7
7.1 Sub-function Hz ...................................................................... 7
8 Current Measurement ................................................8
8.1 Measuring Alternating Current with (Clip-On) Current
Transformers ........................................................................10
8.1.1 Transformer Output mA / A ................................................... 10
8.1.2 Transformer Output V ............................................................10
9 Resistance Measurement .........................................10
10 Continuity and Diode Testing ...................................11
11 Capacitance Measurement ......................................12
12 Frequency Measurement –
Duty Cycle Measurement .........................................12
12.1 Duty Cycle Measurement ....................................................... 12
12.2 RPM Measurement ............................................................... 12
13 Temperature Measurement ......................................13
14 Characteristic Values ...............................................14
15 Maintenance ............................................................17
15.1 Battery ................................................................................. 17
15.2 Fuses ................................................................................... 17
15.3 Housing ................................................................................ 17
16 Recalibration ............................................................18
17 Accessories .............................................................18
17.1 General ................................................................................ 18
17.2 Technical Data for Measurement Cables
(included: KS17-2 safety cable set) ........................................ 18
18 Repair and Replacement Parts Service,
Calibration Center and Rental Instrument Service ...19
19 Manufacturer’s Guarantee .......................................19
20 Product Support .......................................................19
1 Safety Features and Precautions
You have selected an instrument which provides you with a
high level of safety.
This instrument fulfills the requirements of the applicable EU
guidelines and national regulations. We confirm this with the
CE marking. The relevant declaration of conformity can be
obtained from GMC-I Messtechnik GmbH.
The analog-digital multimeter is manufactured and tested in
accordance with safety regulations IEC 61 010–1:2010/
DIN EN 61 010–1:2011/VDE 0411–1:2011. When used for its
intended purpose, safety of the operator, as well as that of
the instrument, is assured. Their safety is however not
guaranteed, if the instrument is used improperly or handled
carelessly.
In order to maintain flawless technical safety conditions, and to
assure safe use, it is imperative that you read the operating
instructions thoroughly and carefully before placing your instrument
into service, and that you follow all instructions contained therein.
In the interest of your own safety and in order to protect the
instrument, the multimeter is equipped with an automatic
socket blocking mechanism. This mechanism is linked to the
function selector switch and only allows access to those
jacks which are actually required for the selected function. It
also prevents the user from turning the selector switch to
impermissible functions after the measurement cables have
already been plugged in.
Observe the following safety precautions:
• The instrument may only be operated by persons who
are capable of recognizing contact hazards and taking
the appropriate safety precautions.
Contact hazards exist
anywhere, where voltages of greater than 33 V
RMS
may occur.
• Avoid working alone when taking measurements which
involve contact hazards. Be certain that a second person
is present.
•Maximum allowable voltage between any of the connector
jacks and earth is 1000 V, category III or 600 V CAT IV.
Nominal voltage at the system may not exceed 600 V.
For the application of measuring cables see chap. 17.2.
Product specificaties
Merk: | Gossen Metrawatt |
Categorie: | Multimeter |
Model: | MetraHit World |
Heb je hulp nodig?
Als je hulp nodig hebt met Gossen Metrawatt MetraHit World stel dan hieronder een vraag en andere gebruikers zullen je antwoorden
Handleiding Multimeter Gossen Metrawatt
19 Maart 2024
27 Februari 2024
27 Februari 2024
27 Februari 2024
27 Februari 2024
27 Februari 2024
27 Februari 2024
27 Februari 2024
9 Juli 2023
8 Juli 2023
Handleiding Multimeter
- Multimeter Abus
- Multimeter Aldi
- Multimeter Alecto
- Multimeter Altai
- Multimeter Amiko
- Multimeter Amprobe
- Multimeter Appa
- Multimeter Basetech
- Multimeter Beha-Amprobe
- Multimeter Benning
- Multimeter Brandson
- Multimeter Brennenstuhl
- Multimeter Bruder Mannesmann
- Multimeter Digitus
- Multimeter Elma
- Multimeter Elro
- Multimeter Extech
- Multimeter Ferm
- Multimeter Flir
- Multimeter Fluke
- Multimeter Gembird
- Multimeter Greenlee
- Multimeter Hager
- Multimeter Hubinont
- Multimeter Ideal
- Multimeter Keithley
- Multimeter Kingcraft
- Multimeter Laserliner
- Multimeter Mastech
- Multimeter Maxwell
- Multimeter Megger
- Multimeter Metrel
- Multimeter Metrix
- Multimeter Micronta
- Multimeter Milwaukee
- Multimeter Monacor
- Multimeter Parkside
- Multimeter PCE Instruments
- Multimeter Perel
- Multimeter Plieger
- Multimeter Powerfix
- Multimeter Profile
- Multimeter Sanwa
- Multimeter Schneider
- Multimeter Silverline
- Multimeter Skandia
- Multimeter Stanley
- Multimeter TFA
- Multimeter Topcraft
- Multimeter Uni-T
- Multimeter Velleman
- Multimeter Voltcraft
- Multimeter Wiha
- Multimeter Workzone
- Multimeter Joy-it
- Multimeter Jung
- Multimeter Clas Ohlson
- Multimeter Steren
- Multimeter Klein Tools
- Multimeter EMOS
- Multimeter Tacklife
- Multimeter Biltema
- Multimeter REV
- Multimeter Etekcity
- Multimeter Testo
- Multimeter Somogyi
- Multimeter Testboy
- Multimeter Yato
- Multimeter Wago
- Multimeter GW Instek
- Multimeter Kyoritsu
- Multimeter Owon
- Multimeter Projecta
- Multimeter Cablexpert
- Multimeter Högert
- Multimeter Multimetrix
- Multimeter Digi-tool
- Multimeter Kewtech
- Multimeter Sonel
- Multimeter Testec
- Multimeter Vimar
- Multimeter Strex
- Multimeter PeakTech
- Multimeter MBS
- Multimeter MGL Avionics
- Multimeter Rigol
- Multimeter Chauvin Arnoux
- Multimeter Aim TTi
- Multimeter CEM
- Multimeter Horex
- Multimeter Cosinus
- Multimeter Kurth Electronic
- Multimeter Weltron
- Multimeter Rohde & Schwarz
- Multimeter EEVBlog
- Multimeter Weidmüller
- Multimeter HT Instruments
- Multimeter Caltek
- Multimeter Elworks
- Multimeter Qian
- Multimeter Bearware
- Multimeter Noyafa
Nieuwste handleidingen voor Multimeter
17 November 2024
14 November 2024
11 November 2024
22 Oktober 2024
16 Oktober 2024
16 Oktober 2024
16 Oktober 2024
16 Oktober 2024
16 Oktober 2024
30 September 2024