Weil Mini Handleiding


Lees hieronder de πŸ“– handleiding in het Nederlandse voor Weil Mini (15 pagina's) in de categorie Niet gecategoriseerd. Deze handleiding was nuttig voor 37 personen en werd door 2 gebruikers gemiddeld met 4.5 sterren beoordeeld

Pagina 1/15
MCC IFU - SPEC1001RevF
Weilβ„’ Mini Chest Compressor (MCC)
INSTRUCTIONS FOR USE
Resuscitation International, LLC
17797 N Perimeter Dr Suite #105
Scottsdale, AZ 85255 USA
Tel: (480) 240 9495
Fax: (480) 419 8144
Email: info@resusintl.com
Website: www.resusintl.com
2
IMPORTANT USER INFORMATION
All users must read and understand the entire instruction for use before operating the miniature chest
compressor (MCC). The purpose of the instructions for use is to explain the use, care, and user
maintenance of the MCC and is not intended to teach cardiopulmonary resuscitation.
Instructions for use should always be easily accessible to the users of MCC.
Required skills
Personnel who intend to use the MCC must be trained in Basic Life Support and/or Advanced Life
Support techniques. Resuscitation International LLC strongly recommends that the MCC only be
operated by: emergency medical technicians, paramedics, nurses, physicians, police, fire rescue
personnel, medical staff, and people who are certified according to the American Heart Association
Guidelines for cardiopulmonary resuscitation, or equivalent. In addition, proper use of the MCC
requires a thorough understanding of the product, appropriate training, and adequate practice with the
device.
Disclaimer
Resuscitation International LLC assumes no responsibility for the use of MCC by personnel that do
not fulfill the requirements listed above.
Resuscitation International LLC does not accept liability for injury to personnel or damage to
equipment that may result from misuse of the MCC.
Under no circumstances shall Resuscitation International LLC be liable for incidental or consequential
damage arising from the use of MCC.
The use of drugs or medical equipment in combination with external chest compressions may reduce
the effectiveness of the compressions. Always refer to the instructions for use for other medical
equipment to assure that the MCC is appropriate to use in conjunction with the mechanical chest
compressions being performed.
Side Effects:
Bruising, injury, vomiting, fractured or broken ribs, and soreness of the chest can be possible side
effects of performing CPRβ‘ . Side effects of CPR are common and considered acceptable
ramifications given that the alternative is clinical death. If the patient is resuscitated, all patients
should be evaluated and assessed for post CPR related injuries.
β‘  Susan Robin (2010, January 24). Side Effects of CPR. Retrieved from http://www.livestrong.com/article/75926-side-effects-cpr/. Summary of
Arkansas State University: Basic Life Support CPR and 123 CPR Inc: Complications of CPR.
3

Federal L
aw res
tricts this device to sale by or on the order of a physician.
 The MCC is intended for use on adults only.
 The MCC is not intended for pregnant patients.
 The current version of the American Heart Association Guidelines does not recommend
the use of mechanical CPR on infants and children.
 The MCC is not intended for patients with traumatic injury (wounds resulting from sudden
physical injury or violence).
 When CPR is indicated, manual CPR should start immediately and should not be
postponed.
 The MCC must be used only in cases where manual CPR would normally be initiated.
 Personnel certified in manual CPR must always be present during the use of the MCC.
 If for any reason the MCC stops operation or mechanical problem occurs, immediately
revert to manual CPR.
 Do not leave the patient or device unattended while the MCC is in operation.
 If the MCC is not positioned correctly in relation to the sternum, there is an increased risk
of damage to the internal organs, ribcage, and circulation may be compromised.
 During transport, regular checks of the position of the MCC on the patient should be
performed. Failure to properly position the MCC may cause injury to the patient.
 Straps or restraints used during transportation must not interfere with the operation of the
MCC.
 Do not block the vent of the MCC.
 Do not submerge the MCC in liquid or operate in water.
 Do not re-use single use disposable components. Reuse of these items, which may have
become contaminated during its first use, may result in subsequent health deterioration
of the patient due to contamination.
 Single use disposable components are not intended for reprocessing or sterilization. This
could pose a risk of harm to the patient.
GENERAL WARNINGS AND CAUTIONS


Product specificaties

Merk: Weil
Categorie: Niet gecategoriseerd
Model: Mini

Heb je hulp nodig?

Als je hulp nodig hebt met Weil Mini stel dan hieronder een vraag en andere gebruikers zullen je antwoorden




Handleiding Niet gecategoriseerd Weil

Weil

Weil Mini Handleiding

22 April 2023

Handleiding Niet gecategoriseerd

Nieuwste handleidingen voor Niet gecategoriseerd