SLOTOSPE016
OPHTHALMOSCOPE
User Manual
DIRECT OPHTHALMOSCOPE
USER MANUAL
An instruction of ophthalmoscope for ophthalmologist. It is
an internally powered device. Please read through this
brochure before use, and keep it handy for future reference.
Manufacturer
SYMBOLS:
Authorized Representative in the
European Community
Symbol for the marking of electrical and
electronics devices according to Directive
2002/96/EC. The device, accessories and the
packaging have to be disposed of waste
correctly at the end of the usage. Please
follow Local Ordinances or Regulations for
disposal.
Type B Applied Part
Attention, consult accompanying doctuments
SAFETY INFORMATION
GENERAL
HOW TO USE THIS DEVICE
MAINTENANCE
PERIODIC SAFETY CHECKS
SPECIFICATION
ELECTRICAL CLASSIFICATION
To assure the correct use of the product, b.c safety measures
should always be followed including the warnings and
cautions listed in this instruction manual.
WARNINGS & CAUTIONS:
• Do not use the device where it may be exposed to
ammable gas.
• Protect this device from water, moisture, Impact,
dropping, direct sunlight, heat and re.
• Do not touch the lens directly by hand or something hard.
• Please use the bulb type specied in this instruction
brochure.
• Keep the device out of the reach of unsupervised infants
and children.
• Do not disassemble or attempt to repair the device or
components.
• Turn on (or off) the device by pushing downwards (or
upwards) as g.1
• Scroll the diopter wheel in two directions as g.2 to select a
suitable diopter, which helps to get a clearer observation
vision for user
• Scroll the diaphragm wheel as g.3 for getting a needed
aperture
• Diagnose by aiming the light to the patient's eye as g.4.
Attention:
it is an intermittent operated device, 1 minute ON, 5 minutes
OFF
1. Changing the bulb ( g.5 )
A. Turn off the power and unscrew the head from the handle
of device
B. Pull the old bulb downwards, and replace with new one
C. Screw the head back on to the handle.
Attention:
• Please use the specialized bulb (Model: 8106) for this
device.
• After the bulb burns out, the temperature will remain high
for approximately 5 minutes. Please wait for a moment to
avoid injury.
2. Changing the batteries ( g.6 )
A. Unscrew the bottom cover from the handle
B. Remove the old AA batteries, and replace with new ones.
Ensure the anode of the batteries points to the head of the
device
C. Screw on the bottom cover of the handle.
Attention:
• AA alkaline batteries are recommended.
• Replace all batteries simultaneously and do not use alkali
manganese batteries with lithium as bulb life will be
reduced.
• Do not leave worn batteries in the handle.
• Please take out batteries from the device if not in use for a
prolonged period.
• Batteries should be disposed of in accordance with
local regulations
Cleaning and and Disinfection:
Clean the surface of device with 75, alcohol solution at least
once a week, avoiding the lenses in the head. And corrosive
or polishing detergent is prohibited.
Attention:
Do not soak, autoclave or let water dip into the device!
Preventive inspection and maintenance to be performed
including the frequency of such maintenance. The following
safety check shold be performed at least every 12 months by
a qualied person who has adequate training, knowledge,
and practical experience to perform these tests.
• Inspect the equipment and accessories for mechanical and
functional damage.
• Inspect the safety relevant labels for legibility.
• Verify that the device functions properly as described in the
instructions for use.
Equiprnent Classication with respect to protection from
electric shock: Internally powered equipment.
Degree of protection from electric shock: TYPE B.
Equipment not suitable for use in the presence of ammable
mixture: Yes.
Mode of operation: intermittent operation.
The manufacturer reserves the right to make technical
changes without notice in the interest of progress. Prior
notices will not be given in case of any amendments within
this manual.
DESCRIPTION:
HS-OP10 ophthalmoscope (hereinafter:device) is a classic
medical device for ophthalmology. This device adopts a
professional lens system, which makes doctors conveniently
get a clear visual eld for examining the interior structures of
eyes, such as dioptricmedia etc.
TECHNICAL SPECIFICATIONS:
• Enhanced XHL Xenon Halogen Technology.
• 8106 bulb used.
• 5 different apertures.
• Range of lenses: -20 D to +20 D
CHANGING SPARE PARTS:
Illumination
Luminance
Direct Illumination
18. 5 Lumen
Bulb Specied Model: &106
2.5V, 0.7A, Φ = 4.8mm
Batteries 2 AA alkaline batteries
Operation
Storage &
Transport
Standards
Temperature 50°F~ 95°F
Humidity 30% ~ 75%
Air pressure 700 hPa - 1060 hPa
Temperature -68°F ~ 158°F
Humidity 10% ~ 90%
(without condensation)
Air pressure 500 hPa - 1060 hPa
MDD 93/42/EEC Class I
IEC 60601-1:1995, IEC 60601-1-2:2007
Product specificaties
Merk: | SereneLife |
Categorie: | Hoortoestel |
Model: | AZSLOTOSPE016 |
Heb je hulp nodig?
Als je hulp nodig hebt met SereneLife AZSLOTOSPE016 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden
Handleiding Hoortoestel SereneLife
8 Juli 2023
1 Juli 2023
30 Juni 2023
23 Juni 2023
22 Juni 2023
Handleiding Hoortoestel
- Hoortoestel Siemens
- Hoortoestel Amplicom
- Hoortoestel Beurer
- Hoortoestel Interton
- Hoortoestel Lanaform
- Hoortoestel Phonak
- Hoortoestel Pyle
- Hoortoestel Starkey
- Hoortoestel Unitron
- Hoortoestel Jabra
- Hoortoestel Oticon
- Hoortoestel Shure
- Hoortoestel Eschenbach
- Hoortoestel Black Wolf
- Hoortoestel CLEAR&WISE
- Hoortoestel Listen Clear
Nieuwste handleidingen voor Hoortoestel
9 April 2024
8 April 2024
8 April 2024
3 April 2024
3 April 2024
20 Februari 2024
23 November 2023
13 Oktober 2023
13 Oktober 2023
11 Augustus 2023