Seaward Y160 Handleiding


Lees hieronder de 📖 handleiding in het Nederlandse voor Seaward Y160 (4 pagina's) in de categorie Niet gecategoriseerd. Deze handleiding was nuttig voor 103 personen en werd door 2 gebruikers gemiddeld met 4.5 sterren beoordeeld

Pagina 1/4
Y160 / Y160H / Y160S
Fault Simulator
Operating Instructions
Bracken Hill, South West Industrial Estate, Peterlee, Co. Durham SR8 2SW. England.
Tel: +44 (0)191- 3511 586
www.seaward.com
Limited Warranty & Limitation of Liability
Seaward guarantees this product for a period of 1 year.
The period of warranty will eective be at the day of delivery.
(c) Copyright 2022
All rights reserved. Nothing from this edition may multiplied, made public any form be or in or
manner, either electronically, mechanically, by photocopying, recording, manner, without or in any
prior written consent from Seaward. This also applies to accompanying drawings diagrams.and
Due a policy continuous development Seaward reserves the right alter the to of to
equipment specication and description outlined this publication without prior notice in
and no of part this publication shall deemed be part of any contract for the be to
equipment unless specically referred inclusion within such contract.to as an
Disposal Old Productof
This product has been designed and manufactured with high quality
materials and components that can be recycled and reused.
When the crossed out wheelie bin symbol attached a product means the product is to it is covered
by the European Directive 2012/19/EU. Please familiarise yourself with the appropriate local
separate collection system for electrical and electronic products.
Please dispose of this product according to local regulations. Do not dispose of this
product along with normal waste material. The correct disposal of this product will help
prevent potential negative consequences for the environment and human health.
User Notes
DO NOT touch the product or test leads while performing checks.
Only perform testing where the Fault Simulator is correctly connected to
the outlet box of the equipment under test, where the simulator is
provided with a plug for connection.
DO NOT plug the fault simulator into any mains outlet
DO NOT exceed stated test voltages
Y160 Instructions for use
The Y160 Fault simulator is intended to provide a quick and simple method of ensuring the test
functions of an Electrical Safety Tester are working as expected. It is assumed the operator is
competent with the operation of the test equipment and is trained to understand the hazards that may
arise with performing High voltage testing.
The Y160 fault simulator series has 3 main test functions, Earth Bond, Insulation Resistance and AC
Dielectric Withstand (Flash) Test.
Note: Y160 is supplied with bare ended connection terminals, and the Y160Hare supplied with a plug
for easy connection to UK output terminals. The Y160S is tted with a Schuko plug for connection or
to equipment with European test sockets.
Notes for HAL Users
For each test below it is advised to create a single test sequence for each test. This way the fail
condition can be logged, Creating all three tests within a single sequence will not be possible, as the
aim is to fail the test. We cannot perform the next test if the sequence fails
Checking the Earth Bond Function
The Y160 is designed to oer a pass and a fail measurement for instruments with a 0.1 limit.
1) Plug the Y160H / Y160S into the test socket of the Test Instrument to be checked.
2) If using a Y160 connect the output clip of the test instrument to the Earth Cable of the
simulator.
3) Connect the Earth Return Clip of the Test Instrument to the external metal connection point.
4) rform an Earth Bond Test for no longer than 5s. Pe
5) Perform the test and observe a reading of approx. 0.1 , if your tester has a limit, set this to
0.09Ohms to initiate a Failure condition.
Checking the Insulation Resistance Function
1) Plug the Y160H / Y160S into the test socket of the Test Instrument to be checked.
2) If using a Y160 ensure the Live and Neutral cables of the fault simulator are connected to the
output terminal of the test equipment, and that the Earth cable is connected to the return
terminal of the test instrument.
3) Perform an Insulation Resistance Test at 500V Test Voltage, a one second test time should be
sucient.
4) Perform the test and observe a reading of 0.4M if performed at ac voltage or 0.2M if Ω Ω
performed at dc
5) If you wish to check the trip circuits, set your lower limit to 1MOhms and ensure the test fails
Checking the Flash Test Function
1) Plug the Y160H / Y160S into the test socket of the Test Instrument to be checked
2) If using a Y160 ensure the Live and Neutral cables of the fault simulator are connected to the
output terminal of the test equipment, and that the Earth cable is connected to the return
terminal of the test instrument.
3) Ensure the Trip Limit of the Flash Tester is set to 5mA.
4) Perform a Flash Test ramping the voltage from 0V, do not exceed 1500V.
5) The unit should produce a fail condition and operate the trip function of the instrument between
0.9 and 1.1kV with a reading greater than 5mA.
Environmental Conditions
The Y160/ Y160H / Y160S ve been designed to perform test and measurements in a dry indoor ha
environment.


Product specificaties

Merk: Seaward
Categorie: Niet gecategoriseerd
Model: Y160

Heb je hulp nodig?

Als je hulp nodig hebt met Seaward Y160 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden




Handleiding Niet gecategoriseerd Seaward

Handleiding Niet gecategoriseerd

Nieuwste handleidingen voor Niet gecategoriseerd