Kyoritsu 8329 Handleiding


Lees hieronder de 📖 handleiding in het Nederlandse voor Kyoritsu 8329 (1 pagina's) in de categorie Niet gecategoriseerd. Deze handleiding was nuttig voor 151 personen en werd door 2 gebruikers gemiddeld met 4.5 sterren beoordeeld

Pagina 1/1
1. Features
This product is an optional accessory and designed for our KEW5050 Ior
leakage current logger. It can supply electrical power to KEW5050 when
combined and used together with MODEL8262 AC adapter and
MODEL7278 earth cable. In this case, KEW5050 can measure voltages
up to 240 V AC (50/ 60 Hz).
2. Safety warnings
This product has been designed, manufactured and tested according to
IEC 61010: Safety requirements for electronic measuring apparatus, and
delivered in the best condition after passing quality control tests. This
instruction manual contains warnings and safety procedures which have
to be observed by the user to ensure safe operation of the product and to
maintain it in safe condition. Therefore, read through these operating
instructions before starting to use the product.
# WARNING
Read through and understand the instructions contained in this manual
before starting to use the product.
Keep the manual at hand to enable quick reference whenever
necessary.
The product is to be used only in its intended applications.
Understand and follow all the safety instructions contained in the
manual.
It is essential that the above instructions are adhered to. Failure to follow
the above instructions may cause injury or product damage, and/or
damage to the equipment under test. Kyoritsu assumes no responsibility
for damage and injury caused by misuse or not following the instructions
in the manual.
The symbol indicated on the product, means that the user must refer to #
the related parts in the manual for safe operation of the product. It is essential
to read the instructions wherever the symbol appears in the manual.
# DANGER: is reserved for conditions and actions that are likely to
cause serious or fatal injury.
# WARNING: is reserved for conditions and actions that can cause
serious or fatal injury.
# CAUTION: is reserved for conditions and actions that can cause
injury or instrument damage.
# DANGER
Do not attempt to use in the presence of flammable or explosive
gasses, or in steam environment.
Never attempt to use the product if its surface or your hand is wet.
The product is to be used only in its intended applications or conditions.
Otherwise, safety functions equipped with the product doesn't work, and
damage to the product or serious personal injury may occur.
When the product is used with KEW5050, the max rated voltage to
ground will be 150 V AC in CAT III or 240 V in CAT II. The rating differs
from the one of reference voltage input terminal of KEW5050. Do not
exceed the measurement category range or not test a circuit in which
higher voltage to ground exists; otherwise, this product, AC adapter or
KEW5050 may be damaged or electrical accident may occur.
Rated power frequency of the product is 50/ 60 Hz.
Use MODEL7273 voltage test leads supplied with KEW5050.
Use MODEL8262 AC adapter supplied with KEW5050.
Ensure that the power plug is firmly connected to the AC adapter.
Always connect the product to KEW5050 first, and then to the
measurement line.
Do not disconnect voltage test leads from the input terminals (N/ L) of
the product during a measurement.
Do not disconnect the product from KEW5050 during a measurement.
Disconnect all the cables connected to the product before replacing
fuses; put a new fuse of the same rating.
# WARNING
Turn off the power switch when connecting to KEW5050.
Keep the fuse holder cover screwed and closed during a measurement.
Never attempt to make measurement if any abnormal conditions are
noted, such as cracks or exposed metal parts.
Do not install substitute parts or make any modification to the product.
Return the product to your local Kyoritsu distributor for repair.
# CAUTION
Examine the conductor under test before starting a measurement; it
might be hot.
Never give strong vibrations or drop shocks.
The product is not dust/ water-proof. Do not use the product at dusty
places or to be spattered.
Care should be taken so as not to damage the cable jacket.
Do not pull or bend the root of the cable to prevent a break in the cable.
Use a damp cloth with neutral detergent or water for cleaning the
product. Do not use abrasives or solvents.
Meaning of symbols on the product:
#User must refer to the explanations in the instruction manual.
Double or reinforced insulation
AC
Crossed-out wheel bin symbol (according to WEEE Directive:
2002/96/EC) indicating that this electrical product may not be
treated as household waste, but that it must be collected and
treated separately.
MODEL 8329
Instruction manual
Power supply adapter
Measurement Category
To ensure safe operation of measuring instruments, IEC 61010 establishes
safety standards for various electrical environments, categorized as O to
CAT IV, and called measurement categories. Higher-numbered categories
correspond to electrical environments with greater momentary energy, so a
measuring instrument designed for CAT III environments can endure
greater momentary energy than one designed for CAT II.
O (None, other)
: Circuits which are not directly connected to the mains
power supply.
CAT II : Electrical circuits of equipment connected to an AC
electrical outlet by a power cord.
CAT III : Primary electrical circuits of the equipment connected
directly to the distribution panel, and feeders from the
distribution panel to outlets.
CAT IV : The circuit from the service drop to the service entrance,
and to the power meter and primary overcurrent protection
device (distribution panel).
3. Specifications
Input voltage 100 to 240 V AC (50 Hz/ 60Hz)
Location for use In-door use, altitude 2000m or less
Operating Temp. &
Hum. range
- 10ºC to 50 ºC, relative humidity 85% or less (no condensation)
Storage Temp. &
Hum. range
- 20ºC to 60ºC, relative humidity 85% or less (no condensation)
Dimension 90.6 (L) x 43 (W) x 47.5 (D) mm * excluding the plug & cable
Weight Approx. 115g
Cable length Approx. 180mm
Fuse rating 500 mA / 600 V AC, fast acting type, Ø6.3 × 32mm
Accessories One fuse (installed), Instruction manual
Applicable
standard
IEC61010-1 CAT III 150 V CAT II 240 V Pollution degree 2
IEC60529 IP 40
4. How to use
Ensure that KEW5050 and MODEL8329 are both powered off.
Connect MODEL7273 voltage test leads to the power input terminals
(N/ L) of MODEL8329 respectively, and then connect the plugs of
MODEL 8329 to the reference voltage input terminals of KEW5050.
Firmly connect the power plug of MODEL8329 to MODEL8262 AC
adapter.
Connect the output terminal of MODEL8262 AC adapter to the female
connector of MODEL7278 earth cable.
Firmly connect the connection terminal of MODEL7278 earth cable to
the AC adapter connector on KEW5050.
Connect the clip of MODEL7278 earth cable to a well-known earth
terminal.
MODEL7278 earth cable
#
DANGER
Always check so as not to connecting the clip other than earth
terminal.
Never connect to a live wire.
Note:
Always earth the product with the supplied earth cable when making
measurements using AC adapter; otherwise, stable readings cannot
be obtained.
Connect the alligator clip of MODEL7273 voltage test lead to the circuit
under test.
Power on KEW5050.
Power on MODEL8329.
Confirm that (AC adapter) icon appears on the LCD of KEW5050.
If (Battery) icon is displayed, the connections may incorrect or the
fuse inside MODEL8329 may have blown.
Reversed procedure is applied to remove the product.
5. Product layout
6. Fuse replacement
Ensure that the power switch is set to OFF.
Disconnect all cables from the product.
Loosen the screw on the top side, and remove the fuse holder cover.
Take out the fuse*.
*500 mA / 600 V AC, fast acting type, Ø6.3 × 32mm
Set a new fuse to the fuse holder.
Attach and screw the fuse holder cover.
DISTRIBUTOR
Kyoritsu reserves the rights to change specifications or designs described in this
manual without notice and without obligations.
11-17 92-2312A
Incoming wire
:Device which is
not directly
connected to the
mains power supply
Socket
Interior wiring
Power plug
Mains input terminal (N/ L)
Power switch
Fuse holder cover
Plug (N(T) / L(R))
Screw
Cable
7278
240V
or less


Product specificaties

Merk: Kyoritsu
Categorie: Niet gecategoriseerd
Model: 8329

Heb je hulp nodig?

Als je hulp nodig hebt met Kyoritsu 8329 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden




Handleiding Niet gecategoriseerd Kyoritsu

Handleiding Niet gecategoriseerd

Nieuwste handleidingen voor Niet gecategoriseerd