Introduction
This service information sheet provides the following information for the T5-600 and T5-1000 Electrical
Testers (hereafter referred to as "the tester").
• Safety information
• Parts and service information
• Specifications
• Cleaning procedure
• Required equipment
• Performance tests
• Parts and accessories list
• Battery replacement procedure
For operating instructions, refer to the T5-600/T5-1000 Electrical Tester Instruction Sheet.
Safety Information
Warning
To avoid possible electric shock or personal injury, follow these guidelines:
• Do not use the tester if it is damaged or operating abnormally. Protection
may be impaired.
• Before each use:
• Make sure the battery door is closed and latched.
• Inspect the tester and test leads. Look for cracks, missing plastic, exposed
metal, or damaged insulation. Replace damaged test leads before using the
tester.
• Verify the tester’s operation by measuring a known voltage.
• Replace the batteries as soon as the low battery indicator (M) appears.
• Do not use the tester around explosive gas, vapor or dust.
• Do not apply more than the rated voltage, as marked on the tester, between
terminals or between any terminal and earth ground.
• Refer servicing to qualified personnel.
• Use caution when working above 30 V ac rms, 42 V ac peak, or 60 V dc.
• When using the probes, keep your fingers behind the finger guards on the
probes.
• Connect the common test lead before you connect the live test lead.
Disconnect the live test lead first.
• The display will not display hazardous voltages in display hold. The voltage
indicator LED ( ) continues to operate.
®
Service Information
Serial Numbers 79559999 and Below
3
General Specifications (continued)
RF Field Specification 0.5 % full scale + (specified accuracy) at 3 V/m
Safety Complies with ANSI/ISA-S82.01-94, UL Classified to IEC 61010, CSA/CAN C22.2 No.1010.1-
92, and EN61010-1 1995.
EMC EN 61326
Certifications In accordance with IEC 61010-1 54CJ
Resolution and Accuracy
Function T5-600 Range T5-1000 Range Resolution Accuracy
K600 V rms 1000 V rms 1 V ±(1.5 % + 2 digits)
L600 V 1000 V 1 V ±(1 % + 1 digit)
?100.0 A 100.0 A 0.1 A ±(3 % + 3 digits)
e1000 1000 1 Ω Ω Ω ±(1 % + 2 digits)
Input Characteristics
Input Protection
Function T5-600 T5-1000
Input impedance (nomial
K600 V rms 1000 V rms 1 MΩ, <100 pF ac-coupled
L600 V rms 1000 V rms 1 MΩ, <100 pF
e600 V rms 1000 V rms
Open Circuit Test Voltage Short Circuit Current
e1.65 V dc (nominal) <600 µA
CAT III protects against transients in a fixed equipment installation such as a distribution panel, and lighting systems in
large buildings.
CAT IV protects against transients from a primary supply such as an electricity meter or an overhead or underground utility
service.
Serial Numbers 79559999 and Below
General Specifications
Calibration One-year calibration cycle.
Maximum Voltage
Between any Terminal
and Earth Ground
T5-600: 600 V rms, Overvoltage Category III
T5-1000: 1000 V rms, Overvoltage Category III
Temperature Operating: -10 ° ° ° ° ° ° ° °C to +50 C (14 F to 122 F); Storage: -30 C to +60 C (-22 F to +140 F)
Altitude Operating: 2000 m (6562 ft); Storage: 10,000 m (32808 ft)
Relative Humidity 0 % to 95 %, 5 ° °C to 30 C (41 C (86 °F to 86°F); 0 % to 75 %, 30 ° °C to 40 °F to 104 °F);
0 % to 45 %, 40 ° °C to 50 C (104 °F to 122 °F)
Battery Type and Life AA (2); 400 hours continuous with alkaline; 200 hours continuous with zinc chloride
Service Information
Equipment Required for Performance Tests
5
Equipment Required for Performance Tests
The following equipment is required for performance tests:
• Fluke 5500A Multi-Product Calibrator, or equivalent
• 0 V to 5 V adjustable dc power supply
• Approximately 8 m (26 ft) of #14 single-conductor magnet wire wound into a butterfly-shaped coil.
Figure 1 shows how to make the coil.
Performance Tests
Use the following procedures to verify the tester’s performance. If the tester fails any of the tests, return it to
Fluke for calibration or repair.
Testing the AC Current Function
The tests in this section require the butterfly coil shown in Figure 1. Note that the 10 conductors in the middle
of the coil cause the tester to read 10 times more current than the calibrator supplies.
1. Put one edge of the butterfly coil in a vice to hold the coil during testing.
2. Set the calibrator to the current and frequency given in Step 1 of Table 1.
3. Connect the butterfly coil to the calibrator.
4. Set the tester to the amperage function.
5. Place the tester’s current fork around the middle of the coil so that the alignment marks are centered and
perpendicular to the wire bundle, as shown in Figure 1.
6. Verify that the tester reads within the display limits shown for Step 1 in Table 1.
7. Apply the currents and verify the tester’s readings for the remaining steps in Table 1.
Table 1. AC Current Tests
Step Calibrator output T5-600/T5-1000 Display Limits
1 10 A, 60 Hz 96.7 to 103.3
2 10 A, 45 Hz 96.7 to 103.3
3 0.5 A, 45 Hz 4.5 to 5.5


Product specificaties

Merk: Fluke
Categorie: Thermometer
Model: T5-600
Gewicht: 300 g
Beeldscherm: LCD
Accu/Batterij voltage: 1.5 V
Waterdicht: Ja
Stofwerend: Ja
DC voltage bereik: 0 - 600 V
Afmetingen (B x D x H): 30.5 x 51 x 203 mm
Type product: Digitale multimeter
AC voltage bereik: 0 - 600 V
AC stroom bereik: 0 - 100 A
Weerstand bereik: 0 - 1000 Ohm

Heb je hulp nodig?

Als je hulp nodig hebt met Fluke T5-600 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden




Handleiding Thermometer Fluke

Fluke

Fluke 566 Handleiding

10 April 2023
Fluke

Fluke 54IIB Handleiding

8 Februari 2023
Fluke

Fluke T5-600 Handleiding

29 Januari 2023
Fluke

Fluke 59 MAX+ Handleiding

27 Januari 2023
Fluke

Fluke VT08 Handleiding

18 Januari 2023
Fluke

Fluke 568 Handleiding

12 Januari 2023
Fluke

Fluke MT4 MAX Handleiding

6 Januari 2023

Handleiding Thermometer

Nieuwste handleidingen voor Thermometer