Low Pressure Pneumatic Test Pump
Instructions
Introduction
The 730LTP40k Low Pressure Pneumatic Test Pump (the Product) is
used to check pressure measurement devices against master reference
gauges. The Product supplies the same pressure to both a reference
gauge and a device under test (DUT) to test the accuracy of the DUT.
Contact Fluke Corporation
Fluke Corporation operates worldwide. For local contact information,
go to our website: www.fluke.com.
To register your product, view, print, or download the latest manual or
manual supplement, go to our website.
Fluke Corporation Fluke Europe B.V.
P.O. Box 9090 P.O. Box 1186
Everett, WA 98206-9090 5602 BD Eindhoven
U.S.A. The Netherlands
+1-425-446-5500
fluke-info@fluke.com
Safety Information
A Warning identifies conditions and actions that pose hazards to the
user; a identifies conditions and actions that may damage the Caution
Product or the equiptment under test.
 Warning
To prevent possible electrical shock, fire, or personal injury and
safe operation and maintenance of the Product:
● Read all safety information before you use the Product.
● Do not alter the Product and use only as specified, or the
protection supplied by the Product can be compromised.
● Use only specified replacement parts.
● Have an approved technician repair the Product.
● Do not use the Product if it is altered or damaged.
● Do not use the Product with flammable, explosive, or poisonous
gasses.
● Do not use the Product above its rated pressure.
● Depressurize the system before removing any hoses, tubing,
devices under test, or accessories from the Product.
● Use filters if impurities from source fluids are likely to
contaminate the Product.
● Clean the Product with mild soap and water if contaminated.
Dry parts and apply manufacturer-specified lubricants, seals,
and gaskets during reassembly.
Table 1 lists the symbols used on the Product or in this document.
Table 1. Symbols
Symbol Description
ο‚»Consult user documentation.
WARNING. RISK OF DANGER.
PN 4940426 October 2020
Β© 2020 Fluke Corporation. All rights reserved.
Specifications are subject to change without notice.
All product names are trademarks of their respective companies.
Product Features
Table 2 shows the features of the Product.
Table 2. Product Features
2
4
1
x 10
x 2
6
7
5
3
Item Description Function
APressure wheel Increases and decreases pressure
in large increments
BPressure release valve Releases pressure
COutput port 1 Connection port for the reference
gauge or the DUT
DFine adjustment wheel Increases and decreases pressure
in small increments
EOutput port 2 Connection port for the reference
gauge or the DUT
FO rings
(7.8 mm x 3.1 mm)
Improves the seal between the
reference gauge/DUT and the
pump
GBulkhead plugs
(M20 x 1.5) Seals pressure in output ports
Operation
Zero the Product before you do a pressure test.
Zero the Product
To zero the Product:
1. Turn , pressure release valve, counterclockwise until the valve is fully B
open.
2. Put one O ring in each of the output ports.
3. Connect the reference gauge to C E, output port 1, or , output port 2.
4. Connect the DUT to the open output port.
5. Select a pressure method to test:
●Positive pressure:
Fully turn A, pressure wheel, counterclockwise to decrease
pressure.
●Vacuum pressure:
Fully turn A, pressure wheel, clockwise to increase pressure.
6. Turn D, fine adjustment wheel, to the middle position.
7. Read the zero pressure from the reference gauge.
8. Turn B, the pressure release valve, clockwise until the valve is fully
closed.
Ascending Pressure
To check ascending pressure:
1. Zero the Product.
2. Select a pressure method to test:
●Positive pressure:
To generate pressure, turn A, pressure wheel, clockwise until the
pressure gets close to the target pressure.
●Vacuum pressure:
To generate pressure, turn A, pressure wheel, counterclockwise
until the pressure gets close to to the target pressure.
3. To get to the target pressure, turn D, fine adjustment wheel, to
increase or decrease pressure as needed.
4. Record the pressure from the DUT.
5. Repeat for each target pressure point up to the full scale.
6. Check descending pressure if needed.
Descending Pressure
To check descending pressure:
1. Zero the Product.
2. Generate ascending pressure greater than the target pressure or up to
the full scale.
3. Select a pressure method to test:
●Positive pressure:
To release pressure, turn A, pressure wheel, counterclockwise until
the pressure gets close to to the target pressure.
●Vacuum pressure:
To release pressure, turn A, pressure wheel, clockwise until the
pressure gets close to to the target pressure.
4. To get to the target pressure, turn D, fine adjustment wheel, to
increase or decrease pressure as needed.
5. Record the pressure from the DUT.
6. Repeat for each target pressure point down to zero.
Troubleshoot the Product
If the Product does not function as expected, try the steps in Table 3.
Table 3. Troubleshoot the Product
Problem SolutionCause
Difficult to maintain
pressure (leakage)
The pressure release valve is open. Close the pressure release valve.
The reference gauge is loose. Tighten the reference gauge.
The DUT is loose. Tighten the DUT.
An O ring is worn. Replace the O ring.
The threads on the Product, the reference gauge, or the
DUT are not smooth.
Apply thread seal tape to the threads of the reference gauge and the
threads of the DUT.
An impurity blocks the circulation of air. Repeatedly apply pressure and suddenly release the pressure until the
impurity comes out.
The DUT connector is too small for the output port of
the Product. Use an adaptor to connect the DUT to the Product.
Difficult to generate
pressure
The pressure release valve is open. Close the pressure release valve.
The reference gauge is loose. Tighten the reference gauge.
The DUT is loose. Tighten the DUT.
An O ring is worn. Replace the O ring.
There is a leak in the DUT. Test another DUT.
Valves or wheels do
not turn freely.
The valve or wheel is turned too tightly. Firmly grasp the valve or wheel and turn it gently.
The threads on the valves or wheels are dry. Apply grease to the threads on the valves and wheels.
Specifications
Operating
Altitude m ...................................... ≀3000
Temperature............................. 23 Β±5 Β°C
Pressure
Media Air .........................................
Positive Range......................... 0 kPa to 40 kPa, 0 psi to 5.8 psi
Vacuum Range -40 kPa to 0 kPa, -5.8 psi to 0 psi ........................
Size (H x W x L) ............................ 240 mm x 140 mm x 130 mm
Weight 1.5 kg ............................................
Safety IEC 61010-1: Pollution Degree 2 ..............................................
Note
The vacuum pressure can reach up to 40 % of the local atmospheric pressure.
LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for two
years from the date of purchase. This warranty does not cover fuses, disposable
batteries, or damage from accident, neglect, misuse, alteration, contamination,
or abnormal conditions of operation or handling. Resellers are not authorized
to extend any other warranty on Fluke’s behalf. To obtain service during the
warranty period, contact your nearest Fluke authorized service center to obtain
return authorization information, then send the product to that Service Center
with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS
FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS
NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL
DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some
states or countries do not allow the exclusion or limitation of an implied
warranty or of incidental or consequential damages, this limitation of liability
may not apply to you.


Product specificaties

Merk: Fluke
Categorie: Pomp
Model: 730LTP40K

Heb je hulp nodig?

Als je hulp nodig hebt met Fluke 730LTP40K stel dan hieronder een vraag en andere gebruikers zullen je antwoorden




Handleiding Pomp Fluke

Handleiding Pomp

Nieuwste handleidingen voor Pomp

Sauermann

Sauermann Omega Handleiding

13 November 2024
Sauermann

Sauermann Delta Handleiding

12 November 2024
Testo

Testo 565i Handleiding

12 November 2024
Simpson

Simpson SCP71008 Handleiding

11 November 2024
MSW

MSW -OP-56-DC Handleiding

5 November 2024
MSW

MSW -FM-N-53 Handleiding

4 November 2024
MSW

MSW -AOP40S Handleiding

15 Oktober 2024
MSW

MSW -MN34M Handleiding

15 Oktober 2024
MSW

MSW -MN34 Handleiding

15 Oktober 2024