MagellanTM 83/8400 Calibration Process Quick Reference Sheet
1 2 3 4
5 6 7 8
Ensure that the Scanner-
Scale is powered-on (the
green top LED and orange
>0< button will be lit or
flashing).
Return the wire lift handle
to its down position and
replace the Top Platter. (If
the Top Platter does not
seat correctly, recheck to
see that the lift handle is
fully retracted.)
After ensuring the platter
is fully seated, press &
release the flashing >0<
button once. Wait for the
scanner to sound two
beeps. (The >0< button
will continue to flash.)
Place 20 lbs. of certified
weights on the platter and
press & release the >0<
button once. The scanner
will emit a trill sound upon
registering the weight.
Remove the weights from
the platter and press &
release the >0< button once
to complete the calibration
sequence. The scanner will
reset and the >0< button
will light steadily.
Remove the Top Platter
from the scanner-scale,
and locate the switch
door.
Open the calibration door
cover by pushing back
towards the optics
chassis. (It may be
necessary to push the
wire lift handle into the
upright position to access
the cover.)
Use your finger to reach
inside the calibration bay,
locate the small plunger
switch and press &
release it once. The
scanner will issue seven
rapid beeps and the >0<
button will begin to flash.
www. datalogic.com © 2013 Datalogic, Inc. • All rights reserved.
Datalogic and the Datalogic logo are registered trademarks of Datalogic S.p.A. in many countries, including the U.S.A. and the E.U.
9Remove the platter and
push the wire lift handle
to the upright position in
preparation for sealing the
calibration door.
10 Fully close the calibration
door and apply the
calibration seal in accor-
dance with local require-
ments. The door is fully
closed when the holes in
the door and the metal
tab are aligned. (See
Sealing Notes.)
No portion of the seal,
whether wire, paper
label or other method,
should be in contact
with the weigh frame
(shown in green) or
scale movement will be
compomised.
Lift Handle Sealing Notes
Conditions to Avoid:
Wire loop too long (not snug)
Calibration door not closed
Seal in contact with weigh frame
CALL DATALOGIC TECHNICAL SUPPORT: 1-800-547-2507
VIEW THE CALIBRATION VIDEO AT: http://www.youtube.com/watch?v=wHYjcTdd9Gk
IF YOU ENCOUNTER PROBLEMS OR HAVE QUESTIONS ABOUT
ANY PART OF THE CALIBRATION PROCESS:
820061014 (Rev. A)October 2013
Product specificaties
Merk: | Datalogic |
Categorie: | Barcode scanner |
Model: | 8400 |
Heb je hulp nodig?
Als je hulp nodig hebt met Datalogic 8400 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden
Handleiding Barcode scanner Datalogic
22 Maart 2023
22 Maart 2023
22 Maart 2023
22 Maart 2023
22 Maart 2023
22 Maart 2023
22 Maart 2023
22 Maart 2023
22 Maart 2023
22 Maart 2023
Andere handleidingen Datalogic
26 Juni 2024
26 Juni 2024
8 Juni 2023
5 Juni 2023
30 Mei 2023
13 Mei 2023
11 Mei 2023
5 Mei 2023
29 April 2023