Appa 61 Handleiding


Lees hieronder de 📖 handleiding in het Nederlandse voor Appa 61 (3 pagina's) in de categorie Multimeter. Deze handleiding was nuttig voor 61 personen en werd door 2 gebruikers gemiddeld met 4.5 sterren beoordeeld

Pagina 1/3
Safety Information
Understand and follow operating instructions carefully.
Use the meter only as specified in this manual; other-
wise, the protection provided by the meter may be
impaired.
WARNING
˙ When using test leads or probes, keep your fingers
behind the finger guards.
˙ Remove test lead from Meter before opening the
battery door or Meter case.
˙ Use the Meter only as specified in this manual or the
protection by the Meter might be impaired.
˙ Always use proper terminals, switch position, and
range for measurements.
˙ Never attempt a voltage measurement with the test
lead inserted into the A input terminal.
˙ Verify the Meter’s operation by measuring a known
voltage. If in doubt, have the Meter serviced.
˙ Do not apply more than the rated voltage, as
marked on Meter, between terminals or between
any terminal and earth ground.
˙ Do not attempt a current measurement when the
open voltage is above the fuse protection rating.
Suspected open circuit voltage can be checked with
voltage function.
˙ Only replace the blown fuse with the proper rating
as specified in this manual.
˙ Use caution with voltages above 30 Vac rms, 42
Vac peak , or 60 Vdc. These voltages pose a shock
hazard.
˙ To avoid false readings that can lead to electric
shock and injury, replace battery as soon as low
battery indicator
<
<
<
<<appears.
˙ Disconnect circuit power and discharge all high-
voltage capacitors before testing resistance,
continuity, diodes, or capacitance.
˙ Do not use Meter around explosive gas or vapor.
˙ To reduce the risk of fire or electric shock do not
expose this product to rain or moisture.
Symbols as marked on the meter and
Instruction card
Risk of electric shock
See instruction card
DC measurement
Equipment protected by double or reinforced
insulation
<
<
<
<< Battery
Fuse
Earth
<AC measurement
Conforms to EU directives
Do not discard this product or throw away
Measuring AC/DC Voltage And
Frequency
Warning
Warning
Warning
WarningWarning
. . . . . .
. . . . . .
. . . . . .
. . . . . .. . . . . . . . . .
. . . .
. . . .
. . . . . . . .
. . . .
. . . .
. . . .
. . . . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .. . . . . .
. . . .
. . . .
. . . .
. . . . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .. . . . . .
. . . .
. . . .
. . . .
. . . . . . . .
. . . .
. . . .
. . . .
. . . . . . . .
Safety sheet
Safety sheet
Safety sheet
Safety sheetSafety sheet
. . . .
. . . .
. . . .
. . . . . . . .
. . . . .
. . . . .
. . . . .
. . . . . . . . . .
. . . .
. . . .
. . . .
. . . . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .. . . . . .
. . . .
. . . .
. . . .
. . . . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .. . . . . .
. . . .
. . . .
. . . .
. . . . . . . .
. . . .
. . . .
. . . .
. . . .. . . .
. . . .
. . . .
. . . .
. . . . . . . .
. . . . .
. . . . .
. . . . .
. . . . .. . . . .
Read First
Testing for Continuity and Diode
APPA
61/62/62T /62R Instruction
Digital Multimeter Card
Resistance
and Capacitance
(Capacitance for 62, 62T, 62R only)
BlackRedBlackRed
Black
Red
Black
Red
Bad Diode
Good Diode
Note – To improve the measurement accuracy of small
value capacitor, record the reading with the test leads
open then substract the residual capacitance of the
Meter and leads from measurement.
CUNKNOWN = CMEASUREMENT
- CRESIDUAL
MIN MAX Record (For 62, 62T, 62R only)
Caution
˙ Disconnect the test leads from the test points before
changing the position of the function rotary switch.
˙
Never connect a source of voltage with the function
rotary switch in / ;/ A // Hz position.
˙ Do not expose Meter to extremes in temperature or
high humidity.
˙ Never set the meter in A function to measure
the voltage of a power supply circuit in equipment
that could result in damage the meter and the
equipment under test.
Maintenance
Do not attempt to repair this Meter. It contains no
user-serviceable parts. Repair or servicing should
only be performed by qualified personnel.
Cleaning
Periodically wipe the case with a dry cloth and
detergent. Do not use abrasives or solvents.
Disable Auto Power Off
NCV
1. Volt sense switch will be activated on any function
or at OFF status.
2. Test leads are not used for the Volt sense test.
3. Press the Volt sense button. The display goes
black, a tone sounds and the red LED lights up to
verify the instrument is operational. The Volt sense
button must be held down to detect the presence
of voltage without use of the leads.
4. If a voltage of 50V to 600V (50 to 500Hz) is
detected near the top of meter a continuous tone
sounds and the red LED near the top of meter
Illuminates.
Fuse Replacement (For 62, 62T, 62R only)
Refer to the following figure to replace fuse :
Battery Replacement
Refer to the following figure to replace the batteries :
Caution
˙ Replace the batteries as soon as the low batteries
indicator "<" appears, to avoid false reading.
˙ Batteries 1.5V x 2
Caution
˙ Use only a fuse with the amperage, interrupt,
voltage, and speed rating specified.
˙ Fuse rating : 10A, 500V
Non-Contact Voltage (Volt sense)
Auto Power Off (Battery Saver)
Manual Ranging and Auto Ranging
10 min
Specifications
General Specifications
Display : 2000 counts.
Polarity Indication : Automatic, positive implied,
negative indicated.
Overrange Indication : “OL” or “-OL”.
Batteries
Life : Alkaline 250 hours
Low Batteries Indication : <” is displayed when
the batteries voltage drops below operating voltage.
Auto Power Off : Approx 10 minutes.
Operating Ambient : Non-condensing
50F,
51.8F ~ 86F (
80% R.H) 87.8F ~ 104F (
75% R.H),
105.8F ~ 122F (
45% R.H)
Storage Temperature : -4F to 140F , 0 to 80%
R.H. when battery removed from Meter.
Temperature Coefficient :
0.15 x (Spec.Accy) / F ,< 64.4F or > 82.4F.
Measure : Samples 2 times per second nominal.
Altitude : 6561.7 ft (2000m)
Safety : Complies with EN61010-1, UL61010-1,
IEC 61010-1,
V/
:
CAT.III. 600V, CAT.II. 1000V.
A : CAT.III. 500V (for 62/62T/62R only)
Pollution degree : 2
Power Requirements :
1.5V x 2 IEC LR03, AM4 or AAA size
Dimensions (W x H x D) : 74mm x 156mm x 44mm
Weight : (320g) including battery.
Accessories : Battery (installed), Test leads and
user manual.
Measuring DC / AC Current
(For 62, 62T, 62R only)
Disconnect
Temperature °C ,°F
R
DCV
Display Hold
Resistance
Diode Check and Continuity
Range Accuracy Voltage Burden
200.0 ~ 200.0K
** ±(0.7% + 3 dgt)
2.000M ** ±(1.0% + 3 dgt)
20.00M * ±(1.5% + 3 dgt)
2V max
Range Resolution Accuracy
10 mV ±(1.5% + 5 dgt)*
* For 0.4V ~ 0.8V
Max.Test Current : 1.5mA
Max. Open Circuit Voltage : 2V
Overload Protection : 600V rms.
Open circuit Voltage : -1.3V approx.
* <100 dgt rolling.
* * The minimum LCD reading is 1400 count in
Auto Ranging Mode.
Limited Warranty
This thermometer is warranted to the original purchaser
against defects in material and workmanship for 1 year
from the date of purchase. During this warranty period,
manufacturer will, at its option, replace or repair the
defective unit, subject to verification of the defect or
malfunction. This warranty does not cover disposable
batteries, or damage from abuse, neglect, accident,
unauthorized repair, alteration, contamination, or
abnormal conditions of operation or handling.
Any implied warranties arising out of the sale of this
product, including but not limited to implied warranties
of merchantability and fitness for a particular purpose,
are limited to the above. The manufacturer shall not be
liable for loss of use of the instrument or other incidental
or consequential damages, expenses, or economic loss,
or for any claim or claims for such damage, expense or
economic loss. Some states or countries laws vary, so
the above limitations or exclusions may not apply to you.
APPA TECHNOLOGY CORP.
9F, 119-1 Pao-Zong Rd.,
Shin-Tien, Taipei, 23115, Taiwan.
P.O.Box. 12-24 Shin-Tien, Taiwan.
Tel: +886-2-29178820
Fax: +886-2-29170848
E-mail: info@appatech.com
http: //www.appatech.com
Temperature (°C )
(For 62T only)
Temperature Accuracy Overload
Protection
-20°C ~ 0°C ±(2% + 4°C )
1°C ~ 100°C ±(1% + 3°C )
101°C ~ 500°C ±(2% + 3°C )
501°C ~ 800°C ±(3% + 2°C )
600V rms
Temperature (°F )
(For 62T only)
Temperature Accuracy Overload
Protection
-4°F ~ 32°F ±(2% + 8°F)
33°F ~ 212°F ±(1% + 6°F)
213°F ~ 932°F ±(2% + 6°F)
933°F ~ 1472°F ±(3% + 4°F)
600V rms
Electrical Specifications
Accuracy is ±(% reading + number of digits) at
23°C ± 5°C < 80%RH.
DC / AC Volts
Range AC Accuracy
200.0mV * Unspecified
2.000V * ±(1.5%+5dgt) 50Hz ~ 300Hz
20.00V ~ 200.0V *
±(1.5%+5dgt) 50Hz ~ 500Hz *
750V AC / 1000V DC
DC Accuracy : ±(0.5% + 2dgt)
Over voltage protection : 1000V DC or 750 V ACrms.
Input Impedance : 10M // less than 100pF.
* CMRR / NMRR : (Common Mode Rejection Ratio)
(Normal Mode Rejection Ratio)
VAC : CMRR > 60dB at DC, 50Hz / 60Hz
VDC
: CMRR > 100dB at DC, 50Hz / 60Hz
NMRR > 50dB at DC, 50Hz / 60Hz
AC Conversion Type :
Average sensing rms indication.
AC conversions are ac-coupled, true rms responding,
calibrated to the sine wave input.
* The minimum LCD reading is 1400 count in Auto
Ranging Mode.
Crest Factor : C.F. = Peak / Rms
+ 1.5% addition error for C.F. from 1.4 to 3
+ 3% addition error for C.F. from 3 to 4
DC / AC Current (For 62, 62T, 62R only)
Range DC Accuracy AC Accuracy Voltage
Burden
±(1.0% + 3 dgt)
±(1.5% + 5 dgt)
50Hz ~ 500Hz
*
2V max
2.000A
10.00A **
Overload Protection :
A input : 10A (500V) fast blow fuse
* AC Conversion Type : Conversion type and
additional specification are same as DC/AC Voltage.
Frequency (For 62, 62T, 62R only)
Overload Protection : 600V rms.
Minimum pulse width : >25 ns
Duty cycle limits : >30% and <70%
Capacitance (For 62, 62T, 62R only)
Range Sensitivity Accuracy
2000Hz ~200.0KHz >1.5 Vac rms,
<5 Vac rms Frequency :
0.01%±1digit
2.000MHz ~ 20.00MHz >2 Vac rms,
<5 Vac rms
Range Accuracy
2.000nF ~ 200.0µF
±(1.9% + 8 dgt)
2.000mF *
Overload
Protection
600V rms
* < 10 dgt of reading rolling.
** Ampere Testing Duty Ratio Table
Ampere Testing Time Rest Time
10A 1min 10min
9A 2min 10min
8A 3min 10min
7A 4min 10min
6A 5min 10min
5A Continually N/A


Product specificaties

Merk: Appa
Categorie: Multimeter
Model: 61

Heb je hulp nodig?

Als je hulp nodig hebt met Appa 61 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden




Handleiding Multimeter Appa

Appa

Appa 91 Handleiding

19 Februari 2023
Appa

Appa 62 Handleiding

12 Februari 2023
Appa

Appa 93N Handleiding

29 Januari 2023
Appa

Appa 62R Handleiding

28 Januari 2023
Appa

Appa 62T Handleiding

6 November 2022
Appa

Appa 61 Handleiding

31 Oktober 2022
Appa

Appa 97 Handleiding

30 Oktober 2022
Appa

Appa 607 Handleiding

15 September 2022
Appa

Appa 72 Handleiding

4 Oktober 2022
Appa

Appa 99 II Handleiding

15 September 2022

Handleiding Multimeter

Nieuwste handleidingen voor Multimeter