Victor VC60B Handleiding


Lees hieronder de 📖 handleiding in het Nederlandse voor Victor VC60B (1 pagina's) in de categorie Diversen. Deze handleiding was nuttig voor 117 personen en werd door 2 gebruikers gemiddeld met 4.5 sterren beoordeeld

Pagina 1/1
DIGITAL
DIGITAL
DIGITAL
DIGITAL
MEGOHM
MEGOHM
MEGOHM
MEGOHM
METER
METER
METER
METER
INSTRUMENT
INSTRUMENT
INSTRUMENT
INSTRUMENT
OPERATION
OPERATION
OPERATION
OPERATION
MANUAL
MANUAL
MANUAL
MANUAL
1.
1.
1.
1.
INTRODUCTION
INTRODUCTION
INTRODUCTION
INTRODUCTION
The digit al m eg-ohmmeter instrument use s lower
power consumption , high change rate , inductance
energy - DCV converter, change 9V-volt to
DC250V/500V/1000V; Also, uses digit-bridge
resistance measurement for measur ing electric
insulation resistance. F eatures include:
Easy and correct readout.
Wide measurement range
High stability and reliability
LCD display for low power consumption and
clear readout
Light-weight and compact construction for easy
operation.
Auto power off and asleep mode
It
is suitable for elevator
,
machine equipment,
telecommunications system check work.
2.
2.
2.
2.
FRONT
FRONT
FRONT
FRONT
P ANEL
P ANEL
P ANEL
P ANEL
DESCRIPTION
DESCRIPTION
DESCRIPTION
DESCRIPTION
1. LCD: display measurement data and "M Ω
"
2. POWER Switch: lock up itself
3. RANGE Switch
4. 5. 6. VOLTAGE select switch
7.
Test
key (PUSH): self return
8. High VOLTAGE indication: LED display
9. L: connect test circuit input jack terminal
10. G: shield input jack terminal
11. 12. E2/E1: connect
to
GND of test object
input jack terminal
3.
3.
3.
3.
SPECIFICA TIONS
SPECIFICA TIONS
SPECIFICA TIONS
SPECIFICA TIONS
3-1.
3-1.
3-1.
3-1.
GENERAL
GENERAL
GENERAL
GENERAL
SPECIFICA TIONS
SPECIFICA TIONS
SPECIFICA TIONS
SPECIFICA TIONS
1. Display
:
60x32mm large window LCD with
max . reading of 1999.
2. Over range indication: only the MSD "1"
display.
3. Power supply: Single, standard 9 volt battery,
NEDA 1604IEC6F22
4. Power Consumption: unload consumption is less
than 300mW.
5. Operation environment: Temperature 0 ℃~40 ℃;
humidity 30 % RH~85% RH .
6. Dimension: 80mm(W) x 185mm (L) x45mm (H)
7. W eight: approx. 360g ( including 9V battery)
3-2.
3-2.
3-2.
3-2.
ELECTRICAL
ELECTRICAL
ELECTRICAL
ELECTRICAL
SPECIFICA TIONS
SPECIFICA TIONS
SPECIFICA TIONS
SPECIFICA TIONS
Note: Median resistance ensures that two terminal voltage
i s not less than 90% of test voltage of low terminal
limit value of measure resistance.
4.
4.
4.
4.
OPERA TION
OPERA TION
OPERA TION
OPERA TION
1.
Press down the
"
POWER" key.
2.
To
select test voltage
(1000V,
500V,
250V)
according
to
need.
3.
To
select RANGE switch according
to
need (see
the above form).
4.
Connect test object electrode
to
input jack terminal
correspondingly.
5.
When measuring cable, connect protection circle
to
"G" jack.
Test
voltage
250 V ±10%
50 0 V ±10%
1000V ±10%
Range
0.1M Ω -20 M Ω
0.1M Ω -50 M Ω
0.1M Ω -100 M Ω
20M Ω -500M Ω
50M Ω -1000 M Ω
10 0M Ω -2000 M Ω
Accuracy
±( 4% of reading + 2d)
Short current
1.7mA
1.7mA
1.4mA
Median
resistance
2M 
2M 
5 M 
Jack position
L, E2
L, E2
L, E1
6. Press down the "PUSH" switch until display value
stability and reading, then release "PUSH".
7. Connect input line "E1 or E2"
to
test object GND
terminal, connect L
to
test circuit terminal; and
asks the "L" connect line
to
hang in the air.
8. If only the figure"1" is displayed
,
over range is
indicated and should read from a higher range .
When RANGE key is in "Down" position ( i.e. in
" "
position)
,
it
means insulation resistance
exceed 2000M Ω
5.
5.
5.
5.
W ARNING
W ARNING
W ARNING
W ARNING
1. When test voltage select key is not press ed down,
it
is possible
to
appear high voltage on output voltage
jack.
2. When measuring
,
firstly, check whether test
voltage select and test voltage remind on LCD are
the same as need voltage.
3.
To
e nsure operation safety, test object must be
remove d from electrified wire netting and short
circuit for fully discharge.
4.
To
e nsure reading is accura
te
,
don't contact test
terminal during measuring.
5. Keep instrument from high temperature position
,
avoid sunlight
to
affect LCD life.
6.
It
is necessary
to
replace battery when a
" "
symbol appears on the LCD display. If store for a
long time, the battery should be taken out.
7. When unload, readings are displayed. T h is is
normal and
it
do es n't affect measure ment .
8. During "M Ω
"
measuring,
it
is possible that
environment interference or insulation material
cause reading unstable.
So
user may connect "G"
terminal
to
test object shield terminal
to
re ading .
9.
To
e nsure safety and decrease interference, us es SI
rubber material measurement line and don't replace
it
as one likes.
6.
6.
6.
6.
ACCESSORIES
ACCESSORIES
ACCESSORIES
ACCESSORIES
1. Measurement cable with clip 1 pair
2. 9V multi-layer battery 1 pc
3. Introduction manual 1 pc
MB-
0060-13


Product specificaties

Merk: Victor
Categorie: Diversen
Model: VC60B

Heb je hulp nodig?

Als je hulp nodig hebt met Victor VC60B stel dan hieronder een vraag en andere gebruikers zullen je antwoorden