DVM860BL VELLEMAN - 1 -
DVM860BL – DIGITAL MULTIMETER
1. Introduction
At the end of its life cycle, dispose of this product in accordance with local and national disposal
regulations. Read the manual thoroughly before bringing this device into service
Thank you for buying the ! Use this device to measure and test: AC & DC voltage & current (also with DVM860BL
optional clamp ), resistance, temperature, batteries, diodes, continuity and transistors. AC97
2. Safety
The is developed in accordance with IEC1010-1 concerning safety requirements for electrical measuring DVM860BL
instruments with an overvoltage category (CAT III) 600V and pollution 2. Respect these safety prescriptions:
• Protect yourself against electroshocks.
• Do not use this device for any other application than those described in this manual.
• Make sure the device was not damaged in transit.
• Make sure the insulation of the test leads is not damaged and/or the wire itself is not exposed.
• Full compliance with safety standards can only be guaranteed if the device is used with the supplied test leads. If
necessary, they should be replaced with identical leads or leads with identical electric ratings. All test leads should
be in good working order.
• Select the right input jack, function and range before use.
• Never exceed the specified limit values for the various measurement ranges.
• Do not touch unused terminals when the meter is connected to a circuit.
• Place the range selector in the highest position if the value to be measured is unknown beforehand.
• Do not measure voltages > 600V above earth ground.
• Exercise extreme caution when working with voltages in excess of 60Vdc or 30Vrms AC. Keep your fingers behind
the probe barriers while using the device.
• Do not connect the leads to a voltage source while the function switch is in one of the following modes: current,
resistance, temperature, battery, diode, transistor or continuity.
• Disconnect all test leads from the circuit to be tested prior to selecting a different function or range.
• Do not use the meter in dusty environments or in the presence of flammable gases and steam.
• Have the device checked by a qualified technician in case of malfunction.
• Never use the meter if the back panel is not in place and firmly fixed.
• Do not use or store the device in areas exposed to direct sunlight, high temperatures or high humidity.
3. Device Description
a. Front Panel
1. LCD display : allows readout of the measured values
2. ON/OFF button : used to activate / deactivate the device
3. RANGE button : used to select auto or manual range
4. FUNC. button : select AC or DC measurement, select °C or °F
5. Selector : used to select the desired function and range
6. 10A jack : input terminal for max. 10A current
7. INPUT jack : input terminal (exception: 10A)
8. COM jack
9. Backlight button ( ) : to illuminates the LCD for a few seconds
10. DATA HOLD button : Displayed data are held
11. MAX HOLD button : Max. measured value is held
DVM860BL VELLEMAN - 2 -
b. Display Symbols
Safety issue,
consult the manual
AC
Alternating current
DC
Direct current
Double insulation
(Protection class II)
AC or DC
Fuse
Earth ground
/
Battery / Battery low
conforms to EU directives
Diode
Continuity buzzer
Measurement with (optional)
clamp
MAX.H
Max. value is being held
DATA-H
Displayed data are held
AUTO
Automatic range selection
CATIII
Overvoltage category III,
pollution degree 2 (IEC1010-1)
4. Maintenance
• Do not remove the back cover while a voltage is being applied. Only a qualified technician should handle repairs.
•
Always disconnect the test leads from all current sources before opening the meter.
• Replace the batteries when the symbol is displayed.
• Replace blown fuses with an identical one or a fuse with identical voltage and current ratings: F 200mA/250V
(quick-acting).
• Use a damp cloth and a mild detergent to clean the device. Never apply abrasives or solvents to the meter
• Place the power switch in the OFF position when the device is not being used
• Remove the batteries to keep them from leaking if the device is to be stored for a long time.
5. Specifications
Max. accuracy is achieved during a one-year period after calibration. Ideal circumstances require a temperature of 18
to 28°C (64 to 82°F) and a max. relative humidity of 75%.
5.1 General Specifications
Sampling Time ±0.4sec.
Display LCD 16mm
Max. Display 1999 points (3 ½ digits)
Polarity Indication "-" indicates negative polarity
Overrange Indication "OL"
Displayed Measuring Unit Unit of electrical capacity
Battery-Low Indication " "-symbol is displayed
Maximum Altitude 7000ft (2000m)
Max. voltage between terminals and earth 600V DC or AC
Fuse Protection F 200mA/250V (quick-acting)
Automatic Power Off 15 minutes
Power Supply 3 x 1.5V AAA-battery
Operating Temperature 0°C to +40°C (+32 to +104°F)
Storage Temperature -10°C to +50°C (+10 to +122°F)
Dimensions 158 x 74 x 32mm
Weight ±250g (batteries included)
DVM860BL VELLEMAN - 3 -
5.2 Electrical Specifications
5.2.1. DC Voltage
Range Resolution Accuracy
200mV 0.1mV
2V 0.001mV
20V 0.01mV
200V 0.1mV
600V 1V
±0.7% of rdg + 2 digits
Input Impedance : 10M Ω
Overload Protection : 250V DC or rms AC for the 200mV-range, 600V DC or V rms AC for the ranges 2V to 600V
Max. Input Voltage : 600V DC
5.2.2. AC Voltage
Range Resolution Accuracy
200mV 0.1mV
2V 0.001mV
20V 0.01mV
200V 0.1mV
±0.8% of rdg + 3 digits
600V 1V ±1.0% of rdg + 3 digits
Input Impedance : 10M Ω
Overload Protection : 250V DC or rms AC for the 200mV-range, 600V DC or AC rms for the ranges 2V to 600V
Frequency Range : 40 to 400Hz
Response : average response, calibration in rms of a sine wave.
Max. Input Voltage : 600V rms AC
5.2.3. DC Current
Range Resolution Accuracy
200µ A 0.1µ A
2000µ A 1µ A
20.00mA 0.01mA
200.00mA 0.1mA
±1.2% of rdg + 3 digits
2.000A 0.001A
10.00A 0.01A
±2.0% of rdg + 10 digits
Overload Protection : F 200mA/250V fuse (quick-acting) for the µA, mA ranges. The 2A and 10A range are unfused.
Max. Input Current : INPUT jack: 200mA, 10A jack: 10A.
Voltage Drop : 200µV, 20mA, 2A: 20mV, 2000µA, 200mA, 10A-range: 200mV.
5.2.4. AC Current
Range Resolution Accuracy
200µ A 0.1µ A
2000µ A 1µ A
20.00mA 0.01mA
200.00mA 0.1mA
±1.5% of rdg + 5 digits
2.000A 0.001A
10.00A 0.01A
±3.0% of rdg + 10 digits
Product specificaties
Merk: | Velleman |
Categorie: | Multimeter |
Model: | DVM860BL |
Gewicht: | 250 g |
Backlight: | Ja |
Bereik temperatuurmeting: | -20 - 1000 °C |
DC voltage bereik: | 0.2 - 600 V |
Afmetingen (B x D x H): | 158 x 74 x 32 mm |
Data hold: | Ja |
AC voltage bereik: | 0.2 - 600 V |
AC stroom bereik: | 0.0002 - 10 A |
Weerstand bereik: | 200 - 20000000 Ohm |
Continuiteit check: | Ja |
Basische nauwkeurigheid (AC stroom): | ±(1.5%+5d), ±(3.0%+10d) |
Basische nauwkeurigheid (AC voltage): | ±(0.8%+3d), ±(1.0%+3d) |
Basische nauwkeurigheid (resistantie): | ±(1.0%±3d), ±(1.0%±1d), ±(1.0%±5d) |
Basische nauwkeurigheid (DC voltage): | ±(0.7%+2d) |
Basische nauwkeurigheid (DC stroom): | ±(1.2%+3d), ±(2.0%+10d) |
DC stroom bereik: | 0.0002 - 10 A |
Heb je hulp nodig?
Als je hulp nodig hebt met Velleman DVM860BL stel dan hieronder een vraag en andere gebruikers zullen je antwoorden
Handleiding Multimeter Velleman
23 November 2023
26 Juli 2023
16 Juni 2023
22 Mei 2023
27 November 2022
27 November 2022
30 April 2023
24 April 2023
24 April 2023
21 April 2023
Handleiding Multimeter
- Multimeter Abus
- Multimeter Aldi
- Multimeter Alecto
- Multimeter Altai
- Multimeter Amiko
- Multimeter Amprobe
- Multimeter Appa
- Multimeter Basetech
- Multimeter Beha-Amprobe
- Multimeter Benning
- Multimeter Brandson
- Multimeter Brennenstuhl
- Multimeter Bruder Mannesmann
- Multimeter Digitus
- Multimeter Elma
- Multimeter Elro
- Multimeter Extech
- Multimeter Ferm
- Multimeter Flir
- Multimeter Fluke
- Multimeter Gembird
- Multimeter Gossen Metrawatt
- Multimeter Greenlee
- Multimeter Hager
- Multimeter Hubinont
- Multimeter Ideal
- Multimeter Keithley
- Multimeter Kingcraft
- Multimeter Laserliner
- Multimeter Mastech
- Multimeter Maxwell
- Multimeter Megger
- Multimeter Metrel
- Multimeter Metrix
- Multimeter Micronta
- Multimeter Milwaukee
- Multimeter Monacor
- Multimeter Parkside
- Multimeter PCE Instruments
- Multimeter Perel
- Multimeter Plieger
- Multimeter Powerfix
- Multimeter Profile
- Multimeter Sanwa
- Multimeter Schneider
- Multimeter Silverline
- Multimeter Skandia
- Multimeter Stanley
- Multimeter TFA
- Multimeter Topcraft
- Multimeter Uni-T
- Multimeter Voltcraft
- Multimeter Wiha
- Multimeter Workzone
- Multimeter Joy-it
- Multimeter Jung
- Multimeter Clas Ohlson
- Multimeter Steren
- Multimeter Klein Tools
- Multimeter EMOS
- Multimeter Tacklife
- Multimeter Biltema
- Multimeter REV
- Multimeter Etekcity
- Multimeter Testo
- Multimeter Somogyi
- Multimeter Testboy
- Multimeter Yato
- Multimeter Wago
- Multimeter GW Instek
- Multimeter Kyoritsu
- Multimeter Owon
- Multimeter Projecta
- Multimeter Cablexpert
- Multimeter Högert
- Multimeter Multimetrix
- Multimeter Digi-tool
- Multimeter Kewtech
- Multimeter Sonel
- Multimeter Testec
- Multimeter Vimar
- Multimeter Strex
- Multimeter PeakTech
- Multimeter MBS
- Multimeter MGL Avionics
- Multimeter Rigol
- Multimeter Chauvin Arnoux
- Multimeter Aim TTi
- Multimeter CEM
- Multimeter Horex
- Multimeter Cosinus
- Multimeter Kurth Electronic
- Multimeter Weltron
- Multimeter Rohde & Schwarz
- Multimeter EEVBlog
- Multimeter Weidmüller
- Multimeter HT Instruments
- Multimeter Caltek
- Multimeter Elworks
- Multimeter Qian
- Multimeter Bearware
- Multimeter Noyafa
Nieuwste handleidingen voor Multimeter
18 November 2024
18 November 2024
18 November 2024
17 November 2024
17 November 2024
14 November 2024
11 November 2024
22 Oktober 2024
16 Oktober 2024
16 Oktober 2024