TDE Instruments Digalox DPM72-MP Handleiding

TDE Instruments Meetapparatuur Digalox DPM72-MP

Lees hieronder de 📖 handleiding in het Nederlandse voor TDE Instruments Digalox DPM72-MP (8 pagina's) in de categorie Meetapparatuur. Deze handleiding was nuttig voor 40 personen en werd door 2 gebruikers gemiddeld met 4.5 sterren beoordeeld

Pagina 1/8
Digalox DPM72-MP Instruction manual (Rev-2020-09)
Visit to download the software “Digalox Manager”.www.digalox.com
Package contents: Panel meter Digalox DPM72, 5$× jumpers, 1$× mounting bracket
1. Intended Use
Indoor use non condensing, non corrosive.
Panel mounting.
Failure to comply with these instructions will void all guarantee and warranty.
In operation, supply the device preferably via screw terminals with 12 to 24$V
AC/DC. If the device is supplied via USB in operation, USB supply voltage must
be at least 5$V.
2. Safety instructions
Note: The measurement inputs of the device can carry life-threatening
voltages!
When working on the device hazardous voltages must not be connected to
the device!
The device must not be used as the only protective device or protective
shutdown.
Read instruction manual carefully before operating the device!
The device is not intended to protect persons or facilities against harm. Specific
devices must be used to guarantee safety (protection relays, cut-off switches,
etc.).
When connecting switches to the terminals J1-J6, only switches must be used
whose isolation voltage is at least twice the maximum occurring measurement
voltage. For example, when measuring 250$V AC switches must be isolated for
at least 500$V.
Do not open the housing!
Do not use the device in the presence of explosive or flammable substances!
All cables carrying hazardous voltages must be secured with external separators.
----- -----1
3. Description
The multi display with RGB backlight is able to display up to 4 parameters
simultaneously. Thresholds can be associated with individual warning backlight
colours. The unit can be switched between measurement modes via an external
switch. Minimum and maximum values of up to four parameters are recorded and
can be displayed optionally using an external switch. Measured values of one
parameter are recorded over a time span of three minutes up to 14 days. The time
base as well as the display of the measuring history can be activated by an external
switch. The measured values remain stored as long as the device is supplied with
voltage.
The following parameters can be adjusted using the configuration software “Digalox
Manager”: scale endpoint, scale caption, display style (pointer, tachometer, bar
graph, and more), splash image, backlight colour, thresholds for alarm output,
threshold warning colour (light, blink), hysteresis, and more. Recorded measurement
values can be read and a continuous transmission of up to four measuring values
can be enabled. Using the software, values can be viewed and exported as CSV file.
Front
1 Unit or free text
2 Threshold values
3 Upper scale caption
4 Graphical measurement display
5 Digital measurement value
6 Lower scale caption
Back
1 2 optocoupler switch outputs
2 DIP switch for setting measurement
mode and upper scale caption
3 Supply voltage input
4 Jumper pins J1 – J6 for
activating/switching of
measurement mode,
graphical historic data display,
min/max display, auto-scaling
5 USB interface for alternative power
supply and software communication
interface
6 Measurement inputs
----- -----2
4. Electrical Connections
Supply
Digalox
+
12 - 24 V Fuse 250 mA
Supply
Voltage direct
max.
500 V
Ilow Ihigh Uhigh Ulow
+/L -/N
Current direct
I
low
I
high
Load
+/L -/N
max.
10 A
U
high
U
low
FF
10 A
Current and voltage direct
I
low
I
high
Load
+/L -/N
max.
10 A
max.
500 V
U
high
U
low
FF
10 A
Current via current transformer
LoadCurrent
Transformer
Ilow Ihigh Uhigh Ulow
max.
10 A
FF
10 A
+/L -/N
Current via current transformer,
voltage direct
LoadCurrent
Transformer
Ilow Ihigh Uhigh Ulow
max.
10 A
FF
10 A
+/L -/N
max.
500 V
Current on high side,
50 V max!
I
low
I
high
+/L -/N
max.
10 A max.
50 V!
U
high
U
low
FF
10 A
Load
Current on high side, voltage direct,
50 V max!
I
low
I
high
+/L -/N
max.
10 A max.
50 V!
U
high
U
low
FF
10 A
Load
Warning:
The device may only be operated in one of the connection options shown
above!
Current measurement on the high side (between plus and load) only for
voltages up to 508V!
In this case the entire device is on high potential. Particularly to be considered
when connecting switches to J1-6.
----- -----3


Product specificaties

Merk: TDE Instruments
Categorie: Meetapparatuur
Model: Digalox DPM72-MP

Heb je hulp nodig?

Als je hulp nodig hebt met TDE Instruments Digalox DPM72-MP stel dan hieronder een vraag en andere gebruikers zullen je antwoorden




Handleiding Meetapparatuur TDE Instruments

Handleiding Meetapparatuur

Nieuwste handleidingen voor Meetapparatuur