RIGOL User’s Guide
Publication Number: UGE22X00-1110
RP1018H
RP1018H High Voltage Probe
Sept. 2013
© 2013 RIGOL Technologies, Inc. All Rights Reserved.
-6-
1. Specification
Model RP 1018H
Division Ratio 1000:1
Input Resistance 200MΩ
Input Capacitor 1.5 PF
Max. Input: DC+AC peak k18 V CAT II
Max. Input: AC RMS. k12 V CAT II
Max. Loading Current 90uA
Bandwidth DC~150MHz
Rise Time s 2.4 n
Signal / Noise > k 60dB at 1 Hz ; > 50dB at 1MHz
DCV Accuracy ≤3% Full Range
ACV Accuracy k ≤3% at 1 Hz
Temp. Coefficient ≤200 PPM / ℃
Compensation Range ~ 10PF 35PF
Safety -Meets EN61010 031 CAT II
Cable Length ± 2M 0.2M
Operation Temperature Range - 10 ~ 55℃
Humidity ) 85% RH or less (at 35℃
Storage - Temperature Range 20 ~ 70℃
Color: Handel/Body / Ye Black llow
Weight / Volume 460g /80(W) x 80(H) X 320(L) mm
2. Precautions Safety
Thank you for purchasing RP1018H high voltage probe. H igh voltage probe can
prevent high voltage testers from subjecting to unexpected shocks when high doing
voltage measurement . Before using the high voltage probe, the operator must work
read and fully understand the safety rules.
In addition, this high voltage probe only for the person who are trained, experienced,
or otherwise qualified to recognize hazardous situations and who trained in the safety
precautions that necessary to avoid possible injury when using such a device.
-7-
Do not work alone when working with high voltage circuits.
For your own safety, inspect the probes for cracks and frayed or broken leads before
each use. If defects are noted, DO NOT USE the probe.
Hands, shoes, floor and work bench must be dry. Avoid making measurements under
humid, damp or other environmental conditions that might affect the safety of the
measurement situation. If possible, always turn the high voltage source off before
connection or disconnection the probe.
The probe body should be kept clean and free of any conductive contamination. The
probe is for indoor use only.
3. Operation
1) Connect the divider probe common lead (alligator clip) to a good earth ground or
reliable ground.
2) Connect the BNC connector to the BNC input of your oscilloscope.
3) Select the desired range of your oscilloscope.
NOTE
Please turn the high voltage source off before making any connections.
4. Compensation Adjustment
Connect the probe to the oscilloscope, and input a 200Hz square wave to the probe tip.
After that adjust the oscilloscope controls to display a few cycles of the waveform.
Adjust the trimmer located in the BNC plug for a flat topped square wave.
Illustration of 200Hz square wave adjustment


Product specificaties

Merk: Rigol
Categorie: Niet gecategoriseerd
Model: RP1018H

Heb je hulp nodig?

Als je hulp nodig hebt met Rigol RP1018H stel dan hieronder een vraag en andere gebruikers zullen je antwoorden




Handleiding Niet gecategoriseerd Rigol

Rigol

Rigol NFP-3 Handleiding

4 September 2024
Rigol

Rigol PVA7250 Handleiding

7 Augustus 2024
Rigol

Rigol PHA1150 Handleiding

7 Augustus 2024
Rigol

Rigol DP811 Handleiding

7 Augustus 2024
Rigol

Rigol RM-DG Handleiding

11 Januari 2024
Rigol

Rigol PVP2150 Handleiding

9 Januari 2024
Rigol

Rigol PVP3150 Handleiding

27 November 2023
Rigol

Rigol PHA2150 Handleiding

27 November 2023
Rigol

Rigol RP7150S Handleiding

27 November 2023

Handleiding Niet gecategoriseerd

Nieuwste handleidingen voor Niet gecategoriseerd