i800
AC Current Clamp
Instruction Sheet
Introduction
The i800 AC Current Clamp, the Clamp, has been designed for use
with multimeters, recorders, power analyzers, safety testers, etc., for
accurate non-intrusive ac current measurement.
Using the latest transformer technology, the Clamp measures
currents from 100 mA to 800 A.
The Clamp complies with the safety standards EN61010-2-032 for
input and output circuits.
Symbols
The table below lists the symbols used on the Clamp and/or in this
manual.
Symbol Description
~ Do not dispose of this product as unsorted municipal
waste. Go to Fluke’s website for recycling information.
W Important Information. See manual.
X Hazardous Voltage. Risk of electric shock.
T Double insulation.
, Application around and removal from HAZARDOUS LIVE
conductors is permissible.
) Conforms to Canadian Standards Association.
P Complies with the relevant European standards.
; Conforms to Australian standards.
CAT III
Equipment designed to protect against transients in
equipment in fixed-equipment installations, such as
distribution panels, feeders and short branch circuits, and
lighting systems in large buildings.
Safety Instructions
Please read this section carefully. It will make you familiar with the
most important safety instructions for handling the Clamp. In this
instruction sheet, a
Warning identifies conditions and actions that
pose hazard(s) to the user. A
Caution identifies conditions and
actions that may damage the Clamp or the test instruments.
WX Warning
The Clamp may only be used and handled by qualified
personnel. To avoid electric shock or personal injury,
follow these precautions:
• Do not use the Clamp if damaged. Always
connect to display device before it is installed
around the conductor.
• If the Clamp is not used as specified by the
manual, protection provided by the Clamp may
be impaired.
• Do not clamp around a conductor before
connecting clamp output to the measuring
equipment.
• Do not disconnect the output from the measuring
equipment while the Clamp is still around a
conductor.
• Connect the Clamp output to measuring
equipment having safety receptacles.
• Do not use prior to consulting local and national
safety codes and use of protective equipment is
recommended.
• Do not hold the Clamp beyond the tactile barrier.
• Before each use, inspect the Clamp. Look for
cracks or missing portions of the housing or
output cable insulation. Also look for loose or
weakened components. Pay particular attention
to the insulation surrounding the jaws. Open and
close the Clamp, noting detent action indicating
proper mechanical operation of SIAC
TM anti short
circuit feature.
• Use caution when working with voltages above
60 V dc, 30 V ac rms or 42 V ac peak. Such
voltages pose a shock hazard.
• Do not use Clamp in wet environments or in
locations that hazardous gases exist.
SIAC™ (Security Internal Anti Short
Circuit)
SIAC™ (Security Internal Anti Short Circuit), patent applied for,
provides protection against short circuits between adjacent primary
conductors when clamped around conductors. This mechanism
ensures that the magnetic circuit separates at the bottom of the jaw
prior to the jaw opening at the top.
A detent action produced by SIAC™ can be felt when opening the
jaws. This detent action should always be present in use. If the jaws
open smoothly, service is indicated and the Clamp should be
inspected at the nearest Fluke Service Center. See “Contacting
Fluke”.
Operating Instructions
WX Warning
To avoid electric shock or personal injury when using
the Clamp:
• Ensure that your fingers are behind the tactile
barrier as shown in Figure 1.
• Do not use the Clamp if any part, including the
lead and connector(s), appears to be damaged or
if a malfunction of the instrument is suspected.
• Clamp around the current carrying conductor
ensuring good contact between the jaw faces.
Center the cable between the Clamp jaws.
Note
The arrow on the labels shows the direction of the current
flow from the source to the load. In order to have a correct
phase relationship between the primary current and the
clamp output, care must be taken regarding the direction
of current flow.
1. Connect the Clamp output connectors between the
common and the appropriate current input jack of a
suitable meter, then switch the meter on.
2. Select a 20 mA ac or greater current range.
3. Clamp the probe around the current-carrying conductor to
be measured (see Figure 1). Make sure the probe jaws
are tightly closed around the conductor.


Product specificaties

Merk: Fluke
Categorie: Niet gecategoriseerd
Model: i800 AC

Heb je hulp nodig?

Als je hulp nodig hebt met Fluke i800 AC stel dan hieronder een vraag en andere gebruikers zullen je antwoorden




Handleiding Niet gecategoriseerd Fluke

Fluke

Fluke ii500 Handleiding

7 Oktober 2024
Fluke

Fluke iSee TC01A Handleiding

9 September 2024
Fluke

Fluke TiX885 Handleiding

7 December 2023
Fluke

Fluke TiX875 Handleiding

7 December 2023
Fluke

Fluke TiX880 Handleiding

7 December 2023
Fluke

Fluke TiX870 Handleiding

7 December 2023
Fluke

Fluke EI-162BN Handleiding

9 Augustus 2023
Fluke

Fluke CNX C3003 Handleiding

8 Augustus 2023
Fluke

Fluke 80i-2010s Handleiding

8 Augustus 2023

Handleiding Niet gecategoriseerd

Nieuwste handleidingen voor Niet gecategoriseerd