Series III Multimeter
Instruction Sheet
Read First: Safety Information
W To avoid damage to the meter, electric shock or injury:
• Never use the meter if the meter or test leads look damaged.
• Be sure the test leads and rotary switch are in the correct
position for the desired measurement.
• Never measure resistance in a circuit when power is applied.
• Never touch the probes to a voltage source when the test
leads are plugged into the 10 A or 300 mA input jack.
• Never use the meter on circuits that exceed 4800 volt-amps.
• Never apply more than the rated voltage between any input
jack and earth ground.
• Be careful when working with voltages above 60 V dc or 30 V
ac rms. Such voltages pose a shock hazard.
• Keep your fingers behind the finger guards on the test probes
when making measurements.
W Warning
To avoid false readings, which could lead to possible
electric shock or personal injury, replace the battery
as soon as the low battery indicator (M) appears.
Symbols
WCaution: Important Information. See instruction sheet.
MAN Manual Range.
TDouble Insulation.
MLow battery. Replace battery.
CAT II Overvoltage Installation Category per IEC 61010-1-95.
CAT II instruments protect against transients from
equipment supplied from the fixed installation, e.g., TV,
PC, and household appliances.
PN 1646063 February 2002 (English)
 1 97, 1 98 00 F uke o p rat on. r ht r se e . r nt d n9 9 , 2 2 l C r o i All ig s e rv d P i e i China.
Terminals
300
mA
10A
FUSED
600V CAT
V
COM
Milliamps
Amps
Volts, Ohms, Diode Test
Common Terminal
hr2f.eps
See Specifications for overload protection.
Pushbutton
Use the pushbutton to either select a fixed range or put the meter
in the Automatic Touch Hold mode.
Autorange
The meter defaults to autorange when first turned on.
Manual Range
Manual ranging is available in V ac, V dc, ohms, A ac, and A dc.
The manual range mode and Touch Hold mode are mutually
exclusive. is displayed when the meter is in the manualMAN
range mode.
Momentary
+_
V
MAN
hr14f.eps
To return to autorange, press for 1 second or change theR
measurement function.
Automatic Touch Hold Mode
W Warning
To avoid electric shock, do not use the Touch Hold
mode to determine if a circuit with high voltage is
dead. The Touch Hold mode will not capture
unstable or noisy readings.
The Touch Hold mode automatically captures and displays stable
readings. Press for 2 seconds. RHOLD is displayed when
the meter is in the Touch Hold mode.
2 seconds
+_
V DC V DC
_+_
single
beep
V V V
HOLD HOLD
V DC
HOLD
hr3f.eps
When the meter captures a new input, it beeps and a new
reading is displayed.
Note
Stray voltages can produce a new reading.
To exit the Touch Hold mode, press R or change the
measurement function. The Touch Hold mode and manual range
mode are mutually exclusive.
Bar Graph
The bar graph shows readings relative to the full scale value of a
displayed measurement range and indicates polarity.
hr16f.eps