Fluke 705 Handleiding


Lees hieronder de 📖 handleiding in het Nederlandse voor Fluke 705 (8 pagina's) in de categorie Niet gecategoriseerd. Deze handleiding was nuttig voor 40 personen en werd door 2 gebruikers gemiddeld met 4.5 sterren beoordeeld

Pagina 1/8
PN 667345 December 1998, Rev.1, 12/99
©1998, 1999 Fluke Corporation, All rights reserved. Printed in U.S.A.
All product names are trademarks of their respective companies. 1
®
Model 705
Loop Calibrator
Calibration Information
W Warning
To avoid electrical shock, remove test leads and any input signals from the
Model 705 Loop Calibrator before opening the case.
Caution
The Model 705 Loop Calibrator contains parts that can be damaged by static
discharge. Follow the standard practices for handling static sensitive
devices.
Introduction
The Calibration Information for the Model 705 Loop
Calibrator (hereafter referred to as the Loop
Calibrator) provides the information necessary to
calibrate and verify the performance of the Loop
Calibrator.
This document includes the following:
• Contacting Fluke for service
• Specifications
• Cleaning
• Replacing the battery
• Disassembly and reassembly
• Recommended test equipment
• Performance verification procedures
• Calibration adjustments
• User-replaceable parts
Complete operating instructions and the warranty
statement for the Model 705 Loop Calibrator are
provided in the that came with it.Instruction Sheet
Contacting Fluke for Service
This Fluke product will be free from defects in
material and workmanship for three years from the
date of purchase.
The complete warranty statement is in the Model 705
Loop Calibrator Instruction Sheet.
To locate an authorized service center, visit us on the
World Wide Web: or call Fluke usingwww.fluke.com
the phone numbers listed below:
• 1-888-99FLUKE (1-800-993-5853) in USA
• 1-800-36-FLUKE (1-800-363-5853) in Canada
• +31 402-678-200 in Europe
• +81-3-3434-0181 in Japan
• +65-738-5655 in Singapore
• +1-425-356-5500 from other countries
Fluke 705
Calibration Information
2
Specifications
Product specifications are provided Tables 1 and 2. These specifications are based on a 1-year calibration cycle and
apply from 18 ºC to 28 ºC (64 ºF to 82 ºF) unless stated otherwise. Accuracy specifications are given as follows:
± ( [ % of reading ] + [ number of least significant digits ] ).
Table 1. General Specifications
Maximum voltage between any
terminal and earth ground or
between any two terminals
30 V
Display: 0 to 28 V with 0.001 V resolution, 0 to 24 mA with 0.001 mA resolution
Operating Temperature: -10 °C to 55 °C
Storage Temperature: -40 ° °C to 60 C
Temperature Coefficient: ± ° ° 0.005 % of range per C for -0 C to 18 C to 55 °C and 28 ° °C
Relative Humidity: 95 % to 30 °C,
75 % to 40 °C,
45 % up to 50 °C,
35 % up to 55 °C
Operationing Altitude: 3000 meters maximum
Power Requirements: Single 9 V battery (ANSI/NEDA 1604A or IEC 6LR61)
Battery Life (Typical): Source Mode: 18 hours; 12 mA into 500 Ω
Measure / Simulate Mode: 50 hours
Shock, Vibration: 1 meter drop test
Random vibration 2 g @ 5 to 500 Hz
Size (H x W x L): 32 mm x 87 mm x 187 mm (1.25 in x 3.41 in x 7.35 in)
Weight: 349 g (12.3 oz). With Holster and flex-stand: 601 g (21.2 oz)
Safety Certified as compliant to CAN/CSA C22.2 No. 1010.1:1992.
Complies with ANSI/ISA S82.01-1994
Certification P and )
Table 2. Accurcy Specifications
Function Range Resolution Accuracy
DC Volts Input + 28 V 0.001 V ± (0.025 % + 1)
DC mA Input 24 mA 0.001 mA ±( 0.02 % + 2)
DC mA Output 0 mA to 24 mA 0.001 mA ± (0.02 % + 2)
E 0 % -100 % - 0 % 40 second smooth ramp
O 0 % -100 % - 0 % 15 second smooth ramp
N 0 % -100 % - 0 % Stair-step ramp in 25 % steps, pausing 5 seconds at each step
Source mode: Compliance: ≥ Ω 1000 at 20 mA
Simulate mode: External loop voltage requirements: 24 V nominal, 30 V maximum, 12 V minimum
Loop Power ≥ 24 V
Percent display - 25 % to 125 %
Input/Output protection Input/Output protection on mA range: resettable, nonreplaceable 0.1 A fuse.
Loop Calibrator
Cleaning
3
Cleaning
W Warning
To avoid electrical shock, remove test
leads and input signals before cleaning.
To clean the case, wipe it with a cloth lightly
dampened with water and a mild detergent.
Do not use abrasives, solvents, or alcohol.
Replacing the Battery
WWarning
To avoid false readings, which could lead
to possible electric shock or personal
injury, replace the battery as soon as the
low battery symbol ( M ) appears on the
Loop Calibrator’s display.
To replace the battery:
1. Lift and remove the battery cover to gain access
the battery compartment. See Figure 1.
2. Lift the battery from the case bottom and insert a
new 9 V battery.
3. Be sure the positive and negative battery posts
are oriented correctly, as noted in the bottom of
the battery compartment.
AAK07F.EPS
Figure 1. Replacing the Battery
Disassembling and Reassembling the
Calibrator
W Warning
To avoid electric shock or personal injury:
• Do not allow water into the case.
• Remove any input signals prior to
removing test leads and opening case.
• When servicing the Loop Calibrator,
use only specified replacement parts.
To disassemble and reassemble the Loop Calibrator,
refer to Figures 2 and 3 as necessary and perform the
following procedure.
Note
Both a Phillips-head screwdriver and a flat-
blade screwdriver are required.
1. To remove the printed circuit assembly (pca):
Remove all test leads from the terminals of the
Loop Calibrator.
2. Set the rotary switch to
OFF.
3. Lift and remove the battery cover to gain access
to the battery compartment. Remove the battery.
4. Remove all Phillips-head screws from the case
bottom.
5. Separate the case top from the case bottom.
Verify that the battery terminals are being guided
through the bottom case before completely
removing the case bottom.
6. Referring to Figure 3, insert a small, flat-blade
screwdriver between the edge of the case top and
the pca. Gently unsnap the case top from the pca.
7. Repeat step 6 for each of the remaining snaps that
secure the pca to the case top.
8. Lift the pca from the case top by its edges.
9. If the elastomeric strips for the keypad assembly
and LCD are stuck to the pca, remove them, but
do not touch the conductive edges.
Note
Before reinserting the pca, make sure that the
rotary knob is in the OFF position.


Product specificaties

Merk: Fluke
Categorie: Niet gecategoriseerd
Model: 705
Soort bediening: Knoppen
Gewicht: 340 g
Breedte: 70 mm
Diepte: 41 mm
Hoogte: 152 mm
Gebruikershandleiding: Ja
Accu/Batterij voltage: 9 V
Temperatuur bij opslag: -40 - 60 °C
Bedrijfstemperatuur (T-T): -10 - 55 °C
Relatieve vochtigheid in bedrijf (V-V): 0 - 90 procent
Overspanningscontrole: Ja
Automatische stap: Ja
Lusspanning: 30 V
Spanningsmeetnauwkeurigheid: 0.25 procent
Spanningsmeetresolutie: 0.001 V
Krokodillenklem: Ja
Inclusief holster: Ja

Heb je hulp nodig?

Als je hulp nodig hebt met Fluke 705 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden




Handleiding Niet gecategoriseerd Fluke

Fluke

Fluke ii500 Handleiding

7 Oktober 2024
Fluke

Fluke iSee TC01A Handleiding

9 September 2024
Fluke

Fluke TiX885 Handleiding

7 December 2023
Fluke

Fluke TiX875 Handleiding

7 December 2023
Fluke

Fluke TiX880 Handleiding

7 December 2023
Fluke

Fluke TiX870 Handleiding

7 December 2023
Fluke

Fluke EI-162BN Handleiding

9 Augustus 2023
Fluke

Fluke CNX C3003 Handleiding

8 Augustus 2023
Fluke

Fluke 80i-2010s Handleiding

8 Augustus 2023

Handleiding Niet gecategoriseerd

Nieuwste handleidingen voor Niet gecategoriseerd

Electrolux

Electrolux STA9 Handleiding

23 November 2024
Electrolux

Electrolux UMP3 Handleiding

23 November 2024
Bosch

Bosch HEZ8TK36UC Handleiding

23 November 2024
Bosch

Bosch HEZ8ZZ36UC Handleiding

23 November 2024
Bosch

Bosch HDZIT301 Handleiding

23 November 2024
Bosch

Bosch HEZBS301 Handleiding

23 November 2024
LG

LG LX-330 Handleiding

23 November 2024
Bosch

Bosch KSZ2AVU00 Handleiding

22 November 2024