Fluke 700P00 Handleiding


Lees hieronder de 📖 handleiding in het Nederlandse voor Fluke 700P00 (4 pagina's) in de categorie Niet gecategoriseerd. Deze handleiding was nuttig voor 50 personen en werd door 2 gebruikers gemiddeld met 4.5 sterren beoordeeld

Pagina 1/4
®
PN 942735 July 1994 Rev.12, 11/01
©1994-2001 Fluke Corporation. All rights reserved. Printed in USA
All product names are trademarks of their respective companies
700 Series Pressure Modules
Instruction Sheet
15 PSID/G
PRESSURE MODULE
700
P
04
1 bar100 kPa
RANGE
Low
High
Dierential
100 PSIG
PRESSURE MODULE
700
P
06
7 bar700 kPa
RANGE
Gage
yz1f.eps
Figure 1.
Introduction
The Fluke 700 Series Pressure Modules allow you to measure
pressure with the Fluke 700 Series Process Calibrators. Read this
sheet before you use the pressure module. This sheet contains
specifications and information about how to avoid damaging the
pressure module. See the Calibrator Users Manual for operating
instructions.
The pressure module measures pressure using an internal
microprocessor. It receives operating power from and sends digital
information to the 700 Series calibrator.
Gage pressure modules have one pressure fitting and measure
pressure with respect to atmospheric pressure. Differential pressure
modules have two pressure fittings and measure the difference
between the applied pressure on the high fitting versus the low fitting.
Figure 1 shows the two types. A differential pressure module functions
like a gage module when the low fitting is open.
Compatibility of Pressure Modules with Fluke
Process Calibrators
Fluke-701, 702, 741, 743, and 744 Documenting Process
Calibrators: The Fluke-700P00 and P27 Pressure Modules are
completely compatible with these calibrators.
Fluke-716, 717, and 718 Pressure Calibrators: The Fluke-700P00
and P27 Pressure Modules will display pressure readings correctly on
716 Pressure Calibrators with firmware V1.2 or higher, or 717 and 718
Pressure Calibrators with firmware V1.3 or higher. To view the software
revision, power-on the calibrator while holding down the MAX key.
Read the REV number in the display. Contact your Fluke Service
Center for a free update of calibrators with lower revision numbers.
Reference PCN 4647.
Fluke-725 Multifunction Process Calibrators: The Fluke-700P00 and
P27 Pressure Modules will display correct pressure readings when
used with Fluke-725 Multifunction Process Calibrators at internal
software revision 1.9 and later. To view the software revision, power-on
the calibrator while holding down the 0% key. Read the REV number in
the display. Contact your Fluke Service Center for a free update of
calibrators with lower revision numbers. Reference PCN 4578.
Box Contents
Pressure module, strap, 1/4 NPT to 1/4 ISO metric adapters, instruction
sheet.
P
T
th
o
A
A
th
is
Table 1. Pressure Module
MODEL RANGE2GAGE OR
DIFFERENTIAL
ISOLATED OR
NONISOLATED
FLUKE-700P00 0 to 1.000 in H
2O
0 to 0.249 kPa
0 to 0.002 bar
Differential Nonisolated
FLUKE-700P01 0 to 10.00 in H
2O
0 to 2.49 kPa
0 to 0.02 bar
Differential Nonisolated
FLUKE-700P02 0 to 1.0000 psi
0 to 6.8900 kPa
0 to 6.89 E2 bar
Differential Nonisolated
FLUKE-700P03 0 to 5.0000 psi
0 to 34.000 kPa
0 to 0.3400 bar
Differential Nonisolated
FLUKE-700P04 0 to 15.000 psi
0 to 100.00 kPa
0 to 1.0000 bar
Differential Nonisolated
FLUKE-700P05 0 to 30.000 psi
0 to 200.00 kPa
0 to 2.0000 bar
Gage Isolated
FLUKE-700P06 0 to 100.00 psi
0 to 700.00 kPa
0 to 7.0000 bar
Gage Isolated
FLUKE-700P27 0 to 300.00 psi
0 to 2000.0 kPa
0 to 20.000 bar
Gage Isolated
FLUKE-700P07 0 to 500.00 psi
0 to 3400.0 kPa
0 to 34.000 bar
Gage Isolated
FLUKE-700P08 0 to 1000.0 psi
0 to 7000.0 kPa
0 to 70.000 bar
Gage Isolated
FLUKE-700P09 0 to 1500.0 psi
0 to 10000.0 kPa
0 to 100.000 bar
Gage Isolated
1. Use of pressure zero function is required to achieve these specifications. Contact your Fl
V1.2 calibrator.
2. Available pressure units (psi, kPa, bar, inHg, mmHg, inH 20, ftH20., kg/cm2, mmH20, cmH2
3. Accuracy specifications apply for 0 to 100% of full scale, 0 to 50°C. 1% FS typical, -10 to
4. Maximum Non-Destructive Pressure: 3X maximum rated pressure, including common mo
5. Maximum Common Mode Pressure: 3X maximum rated pressure. (700P09: 2X, 700P00:
6. Specifications reflect a confidence interval of 95%.
Avoiding Mechanical Damage
To avoid mechanically damaging the pressure module, never apply
more than 10 ft.-lbs. of torque between the pressure module fittings or
between the fittings and the body of the module. Always apply
appropriate torque between the pressure module fitting and connecting
fittings or adapters. Figure 2 shows the correct way and incorrect
ways to use a wrench when applying torque to the pressure module
fitting.
PSD/G
PRESSU RE
MODU L
E
70
0
P
04
ar
yz2f.eps
Figure 2.
Avoiding Corrosion Damage
To avoid damaging the pressure module from corrosion, use it only
with specified media as shown below:
Isolated modules: any medium that is compatible with type 316
stainless steel.
Nonisolated modules: dry, noncorrosive gasses only.
Recommended Measurement Technique
For best results, it is recommended that the module be pressurized to
full scale and then vented to zero pressure (atmosphere) prior to
zeroing and making measurements with the calibrator.
Note
Low range pressure modules may be sensitive to
gravity. For best results, pressure modules 30 psi and
below should be held at the same physical orientation
from the time they are zeroed until the measurement is
complete.


Product specificaties

Merk: Fluke
Categorie: Niet gecategoriseerd
Model: 700P00

Heb je hulp nodig?

Als je hulp nodig hebt met Fluke 700P00 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden




Handleiding Niet gecategoriseerd Fluke

Fluke

Fluke ii500 Handleiding

7 Oktober 2024
Fluke

Fluke iSee TC01A Handleiding

9 September 2024
Fluke

Fluke TiX885 Handleiding

7 December 2023
Fluke

Fluke TiX875 Handleiding

7 December 2023
Fluke

Fluke TiX880 Handleiding

7 December 2023
Fluke

Fluke TiX870 Handleiding

7 December 2023
Fluke

Fluke EI-162BN Handleiding

9 Augustus 2023
Fluke

Fluke CNX C3003 Handleiding

8 Augustus 2023
Fluke

Fluke 80i-2010s Handleiding

8 Augustus 2023

Handleiding Niet gecategoriseerd

Nieuwste handleidingen voor Niet gecategoriseerd