BJZ Industriedienst und Vertrieb, D-75031 Eppingen, Tel: 07262/1064 Fax: 07262/1063 Operating manual WRIST STRAP MONITOR
03/07 - am.doc
Order numbers: C-193 2334 (4,5 mm press stud), C-193 2331(10,3 mm press stud)
The system consists of two parts: monitor and adapter
Mounting and wiring
A 9 V adapter with fixed voltage and a 3,5 mm stereo jack serves as power
supply (external positive pole).
Connect the stereo jack with the monitor and plug the adapter into the power
socket.
It is recommendable to mount the monitor at eye level right above the
workstation by using, for example, Velcro.
The black ground conductor of the monitor must be connected to the protective
conductor of a second power socket by using, for example, a BJZ grounding
box.
Finally, the wrist strap is attached to the monitor.
Function
Place the switch in direction of the word "Aktiv". If the wrist strap is applied
correctly, the "Safe" - LED will be illuminated.
If the red "Fault" - LED is flashing and a warning signal can be heard, this may
have the following reasons:
• the wrist strap has been removed
• there is no connection between the ground and the monitor
• the wrist strap is damaged
• the coil cord is damaged
Identify the reason of the alert and fix the problem.
If the person connected to the monitor wants to leave the workstation, the switch
must be placed in direction of the word "Stand by". In this case the warning
signal is not actuated, but the flashing "Fault" - LED reminds the person to put
on the wrist strap again.
Approach in case of failure
Check the wiring and wrist strap for functional capability. If necessary calibrate
the monitor.
Calibration
1. Necessary tools:
- defined resistance (e.g. resistance box), junction cord
2. Setup
The setup of the components must be arranged like in the following drawing:
Power supply
adapter 9V
Resistance box
Connection to
g
round
Connection for press stud
grounding cord
Borehole for adjustment
Adjustment
When using grounding cords with 1 x 10 6
Ω protective resistance, the resistance
box should be regulated to 3 x 106 .Ω
When using grounding cords with other protective resistance, the resistance box
should be regulated to + 2 x 106 .Ω
Connect the monitor to the power supply and place the switch in direction of
"Aktiv". By means of a screw driver the potentiometer can be reached through
the borehole saying "Adj." Adjust the potentiometer until the LED switches from
red to green, when using the appropriate resistance. If necessary, the borehole
can be sealed after adjustment.