Kern TGD 50-3C Handleiding

Kern Weegschaal TGD 50-3C

Lees hieronder de 📖 handleiding in het Nederlandse voor Kern TGD 50-3C (3 pagina's) in de categorie Weegschaal. Deze handleiding was nuttig voor 24 personen en werd door 2 gebruikers gemiddeld met 4.5 sterren beoordeeld

Pagina 1/3
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
E- -Mail: info@kern sohn.com
Tel.: +49- - -0 [0]7433 9933
Fax: 9933 149 - -+49 [0]7433 -
internet: - www.kern sohn.com
Operating manual
Electronic Carat Pocket Scales
KERN TGD
Version 1.0
2018-06
GB
TGD-BA-e- 1810
TGD- -e- 2BA 1810
GB
KERN TGD
Version 1.0 2018-06
Operating manual
Electronic Carat Pocket Scales
1. Technical data
KERN D - TG 50 3C
TTGD - -A 50 3C
Readability (d) 0.001 g / 0.005 ct
Weighing range (max) 50 g / 250 ct
50 g (F2)
Operating temperature + 10° C .... + 30° C
Humidity of air max 80% air humidity (non -condensing)
Automatic cut after 300 seconds-off
Batteries 4 x 1.5 V AAA (Running time rating 150 hrs)
2. Scope of delivery
•
Pocket scale
•
Weighing pan
•
Power cable
•
Balance cover •
Batteries (4 x 1.5 V AAA) •
Operating manual
3. Keyboard overview
ON/OFF
Switch on/off
MODE Switch- over weighing unit
TARE Taring / Setting to zero
PCS Parts counting
4. Operation
Weighing:
 Place the balance on a firm, level surface. Level balance with foot
screws until the air bubble of the water balance is in the prescribed
circle. Check levelling regularly
 To switch on, press ON/OFF button, the balance completes a self check. As soon as the weight
display appears, the balance is ready for weighing. If required, set to zero by using TARE button.
 Place load on pan and read the weighing result.
 Use the MODE button to switch over into the available weighing settings.
Taring:
 Place the empty weighing container.
 Press the TARE button, the zero display disappears.
 Weigh the material, the net weight will be indicated.
 When the balance is unloaded the saved taring value is displayed with negative sign.
 To delete the stored tare value, remove load from weighing pan and press the TARE button.
3 D- -e- TG BA 1810
Parts counting:
 In weighing mode press PCS button, the balance changes into the counting mode.
 Place the reference
 Keep the PCS been button pressed, the currently set reference quantity is displayed.
 Use the button to set the quantity of the placed reference (selectable 25, 50, 75, 100) and MODE
acknowledge with PCS button.
 Remove reference.
 If necessary, tare the empty balance container.
 Place counted pieces on pan and read the quantity of pieces.
 Use MODE button to return to the weighing mode.
5. Adjustment
 Arrange adjustment weight (see chap. 1).
Info about adjustment weights can be found on the Internet at: http://www.kern-sohn.com
 Ensure that there are no objects on the weighing pan.
 In weighing mode keep the button pressed until MODE <CAL> appears.
 Press the button again and wait until the weight value of the required adjustment weight MODE
will appear flashing.
 Place adjustment weight. Wait until >, followed by the weight value of the placed <PASS
adjustment weight will be displayed. After that the adjustment will be finished successfully.
 In case of a diverging display the adjustment has failed. Repeat adjustment process.
6. Important Notes
• Do not overload balance.
• The occurrence of electromagnetic fields may result in major display deviations. Then change
location.
• Avoid high air humidity, vapours and dust.
• Do not bring the balance directly in contact with liquids.
• Do not leave permanent load on the weighing pan.
• Never operate balance in explosive environment. The serial version is not explosion protected.
• Check the balance regularly with external test weights.
• Warranty claims will be voided when the housing is opened.
7. Error messages
Capacity of batteries exhausted, change batteries
Adjustment required, see chap. 5
The placed weight exceeds the capacity of the balance. Reduce load.
Place the balance on a firm, level surface and level it.
8. Declaration of conformity
The current EC/EU Conformity declaration can be found online in:
www.kern-sohn.com/ce