Kern HMO Handleiding


Lees hieronder de 📖 handleiding in het Nederlandse voor Kern HMO (17 pagina's) in de categorie Meetapparatuur. Deze handleiding was nuttig voor 42 personen en werd door 2 gebruikers gemiddeld met 4.5 sterren beoordeeld

Pagina 1/17
Sauter GmbH
Ziegelei 1
D-72336 Balingen
e-mail: info@kern-sohn.com
Phone : +49-[0]7433- 9933-0
Fax: +49-[0]7433- - 9933 149
Internet: www.sauter.eu
Instruction manual mobile Leeb
hardness tester
SAUTER HMO
Version 2.0
04/2020
GB
HMO- -e- BA 2020
PROFESSIONAL MEASURING
2 HMO- -e- BA 2020
GB
Thank you for purchasing the mobile digital Leeb impact hardness tester from
SAUTER. We hope that you will be very satised with the high quality of this device
and its extensive functionality. For any questions, wishes and suggestions we are at
your disposal.
Table of contents:
1 Before commissioning .................................................................................. 3
2 Introduction .................................................................................................... 3
2.1 Measuring principle .................................................................................................................. 3
2.2 Structure .................................................................................................................................... 4
2.3 Technical data ........................................................................................................................... 5
2.4 Area of application ................................................................................................................... 5
3 Test procedure ............................................................................................... 6
3.1 Preparation of the device 6........................................................................................................
3.2 Preparation of the sample ....................................................................................................... 6
3.3 Test execution .......................................................................................................................... 8
3.4 Browsing ................................................................................................................................... 8
3.5 Print 9............................................................................................................................................
4 Setup ............................................................................................................... 9
4.1 Settings in the main mask ....................................................................................................... 9
4.2 Menu settings ......................................................................................................................... 10
5 Care and maintenance ................................................................................. 14
5.1 Periodic maintenance ............................................................................................................ 14
5.2 Store ......................................................................................................................................... 14
5.3 Calibration ............................................................................................................................... 15
6 Annex ............................................................................................................ 16
6.1 Conversion area ..................................................................................................................... 17
6.2 Scope of delivery .................................................................................................................... 17
HMO- -e- 3 BA 2020
1 Before commissioning
Before putting the device into operation, check the delivery for any transport damage
to the packaging, the plastic case and the device itself. Should this be the case,
SAUTER must be contacted immediately.
2 Introduction
2.1 Measuring principle
The measuring principle for the measurements carried out with this dynamic impact
hardness tester is physically quite uncomplicated. An impact body with a hard metal
tip is hurled against the surface of the test object by the spring force. When the impact
body hits the surface, it can be damaged, which ultimately leads to a loss of kinetic
energy.
This energy loss is calculated on the basis of velocity measurements, at a certain point
in time when the impactor is at a certain distance from the surface, both in the rebound
and impact phases.
The xed magnet in the impact body generates an induction voltage in the simple wire
coil of the impact tester. The voltage of the signal is proportional to the speed of the
impact body. The signal is processed by the electronics so that the hardness value can
be read on the display and then stored.
English
German
Impact phase
Impact phase
Rebound phase
Rebound phase
Time
Time


Product specificaties

Merk: Kern
Categorie: Meetapparatuur
Model: HMO

Heb je hulp nodig?

Als je hulp nodig hebt met Kern HMO stel dan hieronder een vraag en andere gebruikers zullen je antwoorden




Handleiding Meetapparatuur Kern

Kern

Kern ORA 1WB Handleiding

30 November 2024
Kern

Kern HMM Handleiding

10 September 2024
Kern

Kern HDA 100-1 Handleiding

10 September 2024
Kern

Kern HMO Handleiding

10 September 2024
Kern

Kern HO 2K Handleiding

10 September 2024
Kern

Kern HK-D Handleiding

10 September 2024
Kern

Kern HO 5K Handleiding

10 September 2024
Kern

Kern FS 4-500 Handleiding

10 September 2024
Kern

Kern FH 100KUK Handleiding

10 September 2024
Kern

Kern HK-DB Handleiding

10 September 2024

Handleiding Meetapparatuur

Nieuwste handleidingen voor Meetapparatuur