User's Guide
CO2 Monitor

Model CO200

2 CO200-EU-EN v1.2 08/13
Introduction
Congratulations on your purchase of this Model CO200 Carbon Dioxide Meter. This meter
measures CO2 level, air temp., humidity, date and time. With visible and audible alarms, this is an
ideal instrument for indoor air quality (IAQ) diagnosis. This meter is shipped fully tested and
calibrated and, with proper use, will provide years of reliable service.
Meter Description
Meter
1. CO2, Temp and RH sensors (rear)
2. LCD display
3. Controls
LCD Display
1. CO2 concentration in ppm
2. Date and Time
3. Air Temperature
4. % Relative Humidity
Symbols
ppm CO2 value
GOOD CO2 air quality level
NORMAL CO2 l air quality level
POOR CO2 air quality level
Air Temp Air Temperature
Humidity % Relative Humidity
°C or °F Celsius or Fahrenheit
MAX/MIN Maximum or Minimum reading
Relay activated
Controls
SET Enters setup mode.
Saves and finishes settings.
ESC Exits setup page/mode.
Terminates during CO
2
calibration.
RESET Press to clear the MAX/MIN. -
Terminates during RH calibration.
â–² Selects mode or increases value in setup
MIN/MAXâ–¼ Activates MAX, MIN function.
Enters CO
2
calibration with
SET
and
â–²
3 CO200-EU-EN v1.2 08/13
Operation
POWER ON
Plug in the adaptor and the meter turns on automatically with a short beep. If the voltage is
too high or low, “bAt“ will display on the LCD and an LED will flash.
The LCD will display current CO
2
, temp., humidity, date and time. The air quality level is
displayed as well
TAKING MEASUREMENT
The meter starts measurements when powered on and updates readings every second. If the
operating environment changes (ex. from high to low temp.), it takes 30 sec for CO2 sensor to
respond and 30 minutes for RH.
NOTE: Do not hold the meter close to your mouth or any other source of CO2.
MAX/MIN
In the normal mode, press the MIN/MAX button to see the minimum and maximum of CO2,
Temperature and Humidity. Each press of MIN/MAX button will sequence through the MIN, MAX
display and then returns to normal mode.
Press and hold the RESET button for more than 1 second to clear the minimum and maximum
value from memory.
ALARM & OUTPUT
The meter features an audible alarm which will give warnings when the CO2 concentration exceeds
the set limit. There are two settable limits, an upper limit that initiates the alarm and a lower limit to
stop the alarm.
The meter emits beeps and displays the fan icon on the display when the CO2 level goes over the
upper limit. The audible alarm can be stopped by pressing any key or it automatically stops when
CO2 reading falls below the lower limit.
If the beeper is temporarily shut down, it will sound again when readings fall below lower limit and
then go over the upper limit again or it the user presses the RESET button for more than 1 second
to activate it.
The fan icon keeps flashing when beeps are manually shut down. It stops only when readings fall
under the lower limit


Product specificaties

Merk: Extech
Categorie: Meetapparatuur
Model: CO200

Heb je hulp nodig?

Als je hulp nodig hebt met Extech CO200 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden




Handleiding Meetapparatuur Extech

Extech

Extech 380926 Handleiding

22 Oktober 2024
Extech

Extech 380941 Handleiding

22 Oktober 2024
Extech

Extech PRC20 Handleiding

21 Augustus 2024
Extech

Extech PRC30 Handleiding

21 Augustus 2024
Extech

Extech SD900 Handleiding

21 Augustus 2024
Extech

Extech PRC15 Handleiding

21 Augustus 2024
Extech

Extech 412300A Handleiding

21 Augustus 2024
Extech

Extech 412400 Handleiding

21 Augustus 2024
Extech

Extech 433201A Handleiding

21 Augustus 2024
Extech

Extech SD910 Handleiding

21 Augustus 2024

Handleiding Meetapparatuur

Nieuwste handleidingen voor Meetapparatuur