SPECIFICATIONS
AD201
Range 0 to 1999 ppm
Resolution 1 ppm
Accuracy ± 2% full scale
Temperature Automatic
compensation
Conversion factor 0.5 ppm
Environment 0 to 50 °C
95 % Relative
Humidity
Battery Life/Type 150 Hrs./4x1.5 V
Dimensions 173 x 41 x 22 mm
AD202
Range 0.00 to 10.00 ppt
Resolution 0.01 ppt
Accuracy ± 2% full scale
Temperature Automatic
compensation
Conversion factor 0.5 ppm
Environment 0 to 50 °C
95 % Relative
Humidity
Battery Life/Type 150 Hrs./4x1.5 V
Dimensions 173 x 41 x 22 mm
USER MANUAL
AD201 & AD202
TDS Testers
• Direct readings: better resolution
• Automatic temperature compensation
• Easy to read display
Dear Customer,
Thank you for choosing an ADWA product.
Please read carefully this manual before
starting operations.
For additional technical information, please
e-mail us at:
sales@adwainstruments.com
ISTRAD201 09/14
www.adwainstruments.com
WARNING
The enclosed product is intended for use
by persons know-ledgeable in safe
laboratory practices. Failure can result from
surface damage, improper pressure or
temperature, or use with improper chemicals.
Information concerning limitations of
ADWA products can be obtained from
ADWA Kft.
SPECIFICATIONS
AD201
Range 0 to 1999 ppm
Resolution 1 ppm
Accuracy ± 2% full scale
Temperature Automatic
compensation
Conversion factor 0.5 ppm
Environment 0 to 50 °C
95 % Relative
Humidity
Battery Life/Type 150 Hrs./4x1.5 V
Dimensions 173 x 41 x 22 mm
AD202
Range 0.00 to 10.00 ppt
Resolution 0.01 ppt
Accuracy ± 2% full scale
Temperature Automatic
compensation
Conversion factor 0.5 ppm
Environment 0 to 50 °C
95 % Relative
Humidity
Battery Life/Type 150 Hrs./4x1.5 V
Dimensions 173 x 41 x 22 mm
USER MANUAL
AD201 & AD202
TDS Testers
• Direct readings: better resolution
• Automatic temperature compensation
• Easy to read display
Dear Customer,
Thank you for choosing an ADWA product.
Please read carefully this manual before
starting operations.
For additional technical information, please
e-mail us at:
sales@adwainstruments.com
ISTRAD201 09/14
www.adwainstruments.com
WARNING
The enclosed product is intended for use
by persons know-ledgeable in safe
laboratory practices. Failure can result from
surface damage, improper pressure or
temperature, or use with improper chemicals.
Information concerning limitations of
ADWA products can be obtained from
ADWA Kft.
OPERATION
1. Remove the protective cap from the
bottom of the tester.
2. Press ON/OFF button to turn the tester
on.
3. Place the tip of the tester into the sample
to be measured.
4. Record the conductivity value. The tester
will automatically compensate tem-
perature variations.
5. Press ON/OFF button to turn the tester
off.
CALIBRATION
1. Switch the tester on by pressing ON/OFF
button.
2. Place the tip of the probe into the
calibration solution and wait for reading
to stabilize:
- For AD201: use the 1382 ppm TDS
solution.
- For AD202: use the 12.88 mS/cm
conductivity solution and consider a
conversion factor of 0.5: 12.88 mS/cm
conductivity solution corresponds to
6.44 ppt TDS solution.
3. Using the screwdriver supplied, adjust
the calibration trimmer on the back of
the tester to match the calibration
standard value:
- For AD201: 1382 ppm (mg/l)
- For AD202: 6.44 ppt (g/l)
4. The tester is now ready to use.
ALWAYS USE A FRESH SOLUTION
FOR CALIBRATION.
MAINTENANCE
If readings became unstable, clean the
probe by rinsing it in alcohol for ten
minutes. If the readings fades or disappears,
batteries should be replaced.
BATTERY REPLACEMENT
Open the battery compartment at the top
of the tester. Place the batteries noting the
polarity listed in the battery compartment.
OPERATION
1. Remove the protective cap from the
bottom of the tester.
2. Press ON/OFF button to turn the tester
on.
3. Place the tip of the tester into the sample
to be measured.
4. Record the conductivity value. The tester
will automatically compensate tem-
perature variations.
5. Press ON/OFF button to turn the tester
off.
CALIBRATION
1. Switch the tester on by pressing ON/OFF
button.
2. Place the tip of the probe into the
calibration solution and wait for reading
to stabilize:
- For AD201: use the 1382 ppm TDS
solution.
- For AD202: use the 12.88 mS/cm
conductivity solution and consider a
conversion factor of 0.5: 12.88 mS/cm
conductivity solution corresponds to
6.44 ppt TDS solution.
3. Using the screwdriver supplied, adjust
the calibration trimmer on the back of
the tester to match the calibration
standard value:
- For AD201: 1382 ppm (mg/l)
- For AD202: 6.44 ppt (g/l)
4. The tester is now ready to use.
ALWAYS USE A FRESH SOLUTION
FOR CALIBRATION.
MAINTENANCE
If readings became unstable, clean the
probe by rinsing it in alcohol for ten
minutes. If the readings fades or disappears,
batteries should be replaced.
BATTERY REPLACEMENT
Open the battery compartment at the top
of the tester. Place the batteries noting the
polarity listed in the battery compartment.
ACCESSORIES
AD70030P 12.88 mS/cm EC
standard solution, 25 x
20 ml
AD70032P 1382 ppm TDS standard
solution, 25 x 20 ml
ACCESSORIES
AD70030P 12.88 mS/cm EC
standard solution, 25 x
20 ml
AD70032P 1382 ppm TDS standard
solution, 25 x 20 ml